Pediatric Radiology

, Volume 48, Issue 5, pp 715–721 | Cite as

Quantification of scatter radiation from radiographic procedures in a neonatal intensive care unit

  • Mariaconcetta Longo
  • Elisabetta Genovese
  • Salvatore Donatiello
  • Bartolomeo Cassano
  • Teresa Insero
  • Mauro Campoleoni
  • Antonella del Vecchio
  • Andrea Magistrelli
  • Paolo Tomà
  • Vittorio Cannatà
Original Article

Abstract

Background

In a neonatal intensive care unit (NICU), preterm infants are often exposed to a large number of radiographic examinations, which could cause adjacent neonates, family caregivers and staff members to be exposed to a dose amount due to scatter radiation.

Objective

To provide information on scatter radiation exposure levels in a NICU, to compare these values with the effective dose limits established by the European Union and to evaluate the effectiveness of radiation protection devices in this setting.

Materials and methods

Radiation exposure levels due to scatter radiation were estimated by passive detectors (thermoluminescent dosimeters) and direct dosimetric measurements (with a dose rate meter); in the latter case, an angular map of the scatter dose distribution was achieved.

Results

The dose due to scatter radiation to staff in our setting is approximately 160 μSv/year, which is markedly lower than the effective dose limit for workers established by the European Union (20 mSv/year). The doses range between 0.012 and 0.095 μSv/radiograph. Considering a mean hospitalization period of 3 months and our NICU workload, the corresponding scatter radiation dose to an adjacent patient and/or his/her caregiver is at most 40 μSv.

Conclusion

For distances greater than 1 m from the irradiation field, both scatter dose absorbed by a staff member during a year and that by an adjacent patient and/or his/her caregiver during hospitalization is less than 1 mSv, which is the exposure limit for public members in a year.

Keywords

Infants Neonatal intensive care unit Neonates Patient dose Radiation protection Radiation protection devices Scatter radiation 

Notes

Compliance with ethical standards

Conflicts of interest

None

References

  1. 1.
    Alzimami K, Sulieman A, Yousif A et al (2014) Evaluation of radiation dose to neonates in a special care baby unit. Radiat Phys Chem 104:150–153CrossRefGoogle Scholar
  2. 2.
    Smans K, Struelens L, Smet M et al (2008) Patient dose in neonatal units. Radiat Prot Dosim 131:143–147CrossRefGoogle Scholar
  3. 3.
    Suliman II, Elawed SO (2013) Radiation dose measurements for optimization of chest X-ray examinations of children in general radiography hospital. Radiat Prot Dosim 156:310–314CrossRefGoogle Scholar
  4. 4.
    Aramesh M, Zanganeh KA, Dehdashtian M et al (2017) Evaluation of radiation dose received by premature neonates admitted to neonatal intensive care unit. J Clin Med Res 9:124–129CrossRefPubMedGoogle Scholar
  5. 5.
    Robinson A (1983) Radiation doses to neonates requiring intensive care. Br J Radiol 56:397–400CrossRefPubMedGoogle Scholar
  6. 6.
    Frayre AS, Torres P, Gaona E et al (2012) Radiation dose reduction in a neonatal intensive care unit in computed radiography. Appl Radiat Isot 71:57–60CrossRefPubMedGoogle Scholar
  7. 7.
    The Council of the European Union (1997) Council Directive 97/43/Euratom of 30 June 1997 on health protection of individuals against the dangers of ionizing radiation in relation to medical exposure, and repealing Directive 84/466/Euratom. https://publications.europa.eu/en/publication-detail/-/publication/aa7564fa-fd07-4872-943c-66df8f4f1099/language-en. Accessed 22 Jan 2018
  8. 8.
    Yu C (2010) Radiation safety in the neonatal intensive care unit: too little or too much concern? Pediatr Neonatol 51:311–319CrossRefPubMedGoogle Scholar
  9. 9.
    Armpilia C, Fife AJ, Croasdale PL (2002) Radiation dose quantities and risk in neonates in a special care baby unit. Br J Radiol 75:590–595CrossRefPubMedGoogle Scholar
  10. 10.
    Fernandez R, Moreno-Torres M, Contreras AM et al (2015) Patient and staff dosimetry during radiographic procedures in an intensive care unit. J Radiol Prot 35:727–732CrossRefPubMedGoogle Scholar
  11. 11.
    Makri T, Yakoumakis E, Papadopoulou D et al (2006) Radiation risk assessment in neonatal radiographic examinations of the chest and abdomen: a clinical and Monte Carlo dosimetry study. Phys Med Biol 51:5023CrossRefPubMedGoogle Scholar
  12. 12.
    Gialousis GI, Yakoumakis EN, Papadopoulou DI et al (2006) Differences in effective dose and energy imparted estimation from PA-AP, RLAT-LLAT projections in pediatric full spine x-ray examination using the Monte Carlo technique. Phys Med Biol 51:287–297CrossRefPubMedGoogle Scholar
  13. 13.
    Trinh AM, Schoenfeld AH, Levin TL (2010) Scatter radiation from chest radiographs: is there a risk to infants in a typical NICU? Pediatr Radiol 40:704–707CrossRefPubMedGoogle Scholar
  14. 14.
    Duetting T, Foerste B, Knoch T et al (1999) Radiation exposure during chest X-ray examinations in a premature intensive care unit: phantom studies. Pediatr Radiol 29:158–162CrossRefPubMedGoogle Scholar
  15. 15.
    Poznanski AK, Kanellitsas C, Roloff DW et al (1974) Radiation exposure to personnel in a neonatal nursery. Pediatrics 54:139–141PubMedGoogle Scholar
  16. 16.
    Iyer KR, Spitz L, Kiely E et al (1995) Scattered radiation in a neonatal surgical unit. Eur J Pediatr Surg 5:286–287CrossRefPubMedGoogle Scholar
  17. 17.
    Burrage JW, Rampant PL, Beeson BP (2003) Scatter and transmission doses from several pediatric X-ray examinations in a nursery. Pediatr Radiol 33:704–708CrossRefPubMedGoogle Scholar
  18. 18.
    Bacci C, D’ Angelo L, Furetta C et al (1993) Comprehensive study on LiF:Mg,Cu,P (GR-200A). Radiat Prot Dosim 47:215–218CrossRefGoogle Scholar
  19. 19.
    Furetta C, Leroy C, Lamarche F (1994) A precise investigation on the TL behaviour of LiF:Mg, Cu,P (GR-200A). Med Phys 21:1605–1609CrossRefPubMedGoogle Scholar
  20. 20.
    Hall JG, Foster-Iskenius UG, Allanson JE (1989) Handbook of normal physical measurements. Oxford University Press, OxfordGoogle Scholar
  21. 21.
    Giacco G, Cannatà V, Furetta C et al (2001) On the use of pediatric phantoms in the dose evaluation during computed tomography (CT) thorax examinations. Med Phys 28:199–204CrossRefPubMedGoogle Scholar

Copyright information

© Springer-Verlag GmbH Germany, part of Springer Nature 2018

Authors and Affiliations

  • Mariaconcetta Longo
    • 1
  • Elisabetta Genovese
    • 1
  • Salvatore Donatiello
    • 1
  • Bartolomeo Cassano
    • 1
  • Teresa Insero
    • 1
  • Mauro Campoleoni
    • 2
  • Antonella del Vecchio
    • 3
  • Andrea Magistrelli
    • 4
  • Paolo Tomà
    • 4
  • Vittorio Cannatà
    • 1
  1. 1.Medical Physics UnitIRCCS Bambino Gesù Children’s HospitalRomeItaly
  2. 2.Medical Physics UnitFoundation IRCCS Ca’ Granda Ospedale Maggiore PoliclinicoMilanItaly
  3. 3.Department of Medical PhysicsIstituto Scientifico San RaffaeleMilanItaly
  4. 4.Department of Diagnostic ImagingIRCCS Bambino Gesù Children’s HospitalRomeItaly

Personalised recommendations