Improving resolution of head and neck CTA using the small x-ray tube focal spot

  • Johannes RosskopfEmail author
  • Michael Braun
  • Meinrad Beer
  • Bernd L. Schmitz
Short Report
  • 37 Downloads

Abstract

CT-resolution depends on tube focal spot size. To investigate the effect of reducing focal spot on vessel wall clarity, 22 patients underwent head and neck CTA acquisition with large and small focal spot size. Hounsfield density profile was assessed for each head and neck vessel separately, and significantly sharper density increase at vessel borders was observed using smaller tube focus by an average of 9.9% to 82° angle degrees (p < .05). In conclusion, spatial resolution of CTA of head and neck vessels can be improved by using the small x-ray tube focal spot.

Keywords

Computed tomography angiography Radiographic image enhancement Cerebral angiography Arteries Neck 

Abbreviations

CT

Computed tomography

CTA

Computed tomography angiography.

Notes

Acknowledgements

The authors would like to acknowledge the radiographers (Department of Neuroradiology, University of Ulm, Günzburg, Germany) for the MRI data acquisitions.

Funding

No funding was received for this study.

Compliance with ethical standards

Conflict of interest

The authors declare that they have no conflict of interest.

Ethical approval

All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards.

Informed consent

Informed consent was obtained from all individual participants included in the study.

References

  1. 1.
    Perisinakis K, Papadakis AE, Damilakis J (2009) The effect of x-ray beam quality and geometry on radiation utilization efficiency in multidetector CT imaging. Med Phys 36:1258–1266.  https://doi.org/10.1118/1.3095768 CrossRefGoogle Scholar
  2. 2.
    Gill MK, Vijayananthan A, Kumar G, Jayarani K, Ng K-H, Sun Z (2015) Use of 100 kV versus 120 kV in computed tomography pulmonary angiography in the detection of pulmonary embolism: effect on radiation dose and image quality. Quant Imaging Med Surg 5:524–533.  https://doi.org/10.3978/j.issn.2223-4292.2015.04.04 Google Scholar
  3. 3.
    Haidekker MA (2013) Medical imaging technology, 1st edn. Springer, New YorkCrossRefGoogle Scholar
  4. 4.
    Kueh A, Warnett JM, Gibbons GJ, Brettschneider J, Nichols TE, Williams MA, Kendall WS (2016) Modelling the penumbra in computed tomography. J Xray Sci Technol 24:583–597.  https://doi.org/10.3233/XST-160576 Google Scholar
  5. 5.
    Oh LCW, Lau KK, Devapalasundaram A, Buchan K, Ardley N, Huynh M (2014) Efficacy of ‘fine’ focal spot imaging in CT abdominal angiography. Eur Radiol 24:3010–3016.  https://doi.org/10.1007/s00330-014-3368-6 CrossRefGoogle Scholar
  6. 6.
    Behling R (2017) Medical X-ray sources now and for the future. Nucl Instruments Methods Phys Res Sect A Accel Spectrometers, Detect Assoc Equip 873:43–50.  https://doi.org/10.1016/j.nima.2017.05.038 CrossRefGoogle Scholar
  7. 7.
    Kalender WA (2006) Computertomographie: Grundlagen, Gerätetechnologie, Bildqualität, Anwendungen. 2. Auflage Public Verlag, BingenGoogle Scholar
  8. 8.
    Schneider CA, Rasband WS, Eliceiri KW (2012) NIH image to ImageJ: 25 years of image analysis. Nat Methods 9:671–675.  https://doi.org/10.1038/nmeth.2089 CrossRefGoogle Scholar

Copyright information

© Springer-Verlag GmbH Germany, part of Springer Nature 2019

Authors and Affiliations

  1. 1.Department of RadiologyUniversity of UlmUlmGermany
  2. 2.Department of NeuroradiologyUniversity of UlmGünzburgGermany

Personalised recommendations