Analytical and Bioanalytical Chemistry

, Volume 411, Issue 24, pp 6487–6496 | Cite as

A rapid and quantitative technique for assessing IgG monomeric purity, calibrated with the NISTmAb reference material

  • Peter P. Reader
  • Rouslan V. Olkhov
  • Shaun Reeksting
  • Anneke Lubben
  • Christopher J. Hyde
  • Andrew M. ShawEmail author
Open Access
Research Paper

Abstract

The fraction of intact monomer in a sample (moles/moles), the monomeric purity, is measured as a quality control in therapeutic monoclonal antibodies but is often unknown in research samples and remains a major source of variation in quantitative antibody-based techniques such as immunoassay development. Here, we describe a novel multiplex technique for estimating the monomeric purity and antigen affinity of research grade antibody samples. Light scattering was used to simultaneously observe the mass of antibody binding to biosensor surfaces functionalised with antigen (revealing Fab binding kinetics) or protein A/G (PAG). Initial estimates of monomeric purity in 7 antibody samples including a therapeutic infliximab biosimilar were estimated by observing a mass deficit on the PAG surface compared to the NISTmAb standard of high monomeric purity. Monomeric purity estimates were improved in a second step by observing the mass of antigen binding to the mass of antibody on the PAG surface. The NISTmAb and infliximab biosimilar displayed tightly controlled stoichiometries for antigen binding of 1.31 ± 0.57 and 1.71 ± 0.16 (95% confidence interval)—within the theoretical limit of 1–2 antigens per antibody depending on avidity. The other antibodies in the panel displayed antigen binding stoichiometries in the range 0.06–1.15, attributed to lower monomeric purity. The monomeric purity estimates were verified by electrospray ionization mass spectrometry (ESI), the gold standard technique for structural characterization of antibodies. ESI data indicated that the NISTmAb and infliximab biosimilar samples had monomeric purity values of 93.5% and 94.7%, respectively, whilst the research grade samples were significantly lower (54–89%). Our results demonstrate rapid quality control testing for monomeric purity of antibody samples (< 15 min) which could improve the reproducibility of antibody-based experiments.

Keywords

Biosensor Antibody Monomeric purity NISTmAb Immunoassay 

Introduction

In non-therapeutic antibodies used for research, a link has been made between poor reproducibility of results and variability in antibody quality [1, 2, 3]. Antibody samples are known to undergo degradation over time during storage [4] and in vivo and the fraction of intact monomer in a sample, the monomeric purity, is tested to ensure potency of therapeutic antibodies [5]. Antibodies for research should also be assessed for monomeric purity to ensure reproducible experiments [6] and a rapid technique for assessing the monomeric purity of antibody samples could test the extent of degradation in samples immediately before use.

Lot-to-lot comparisons of antibody samples should be performed with reference standards [7]. The National Institute of Standards and Technology (NIST) provides a monoclonal antibody, the NISTmAb, designed to support biopharmaceutical innovation by serving as reference standard for comparable evaluation of antibody-based techniques between different laboratories [8, 9, 10]. The NISTmAb is a recombinant humanized IgG1ĸ with a known sequence [11] specific to the respiratory syncytial virus protein F (RSVF) [12]. The NISTmAb is well characterized with a known monomeric purity variation of 96.7–98.7% between batches, as characterized by comparing the relative peak areas of a non-denaturing size exclusion chromatogram [8, 9]. The NIST 8671 certificate for the material used in the present study quotes a monomeric purity of 96.63 ± 0.46%.

Fragmentation in monoclonal antibody samples is frequently observed at the hinge region disulphide bonds. In this region, a single polypeptide connects the Fab and Fc fragments and hence cleavage is followed by separation of these fragments [13]. The most abundant antibody fragments and their respective masses observed by non-denaturing size exclusion chromatography (SEC) [4] are shown in Fig. 1a.
Fig. 1

Schematic of fragmentation patterns of a typical IgG and their binding properties: a structure of IgG. Variable and constant domains are marked V and C with subscripts for light and heavy chain marked L or H. Heavy chain regions are numbered 1, 2, and 3 starting from the N-terminus and black lines indicate disulphide bonds. The most abundant antibody fragments recorded in IgG samples and their respective molecular masses are also shown. Heavy chain and light chain fragments are labelled Hc and Lc. The 50 kDa fragment may originate from the Fab region as shown or the Fc region. Similarly, the 25 kDa fragment may originate from the Lc as shown, the Hc of the Fab region, or Hc of the Fc region. b Biophotonic assays used to characterize the monomeric purity of an IgG sample. The Fab assay uses a sensor surface functionalised with antigen (orange) to confirm antibody specificity and derive antibody-antigen binding kinetics. The Fc assay uses a sensor surface functionalised with PAG (blue) to capture Fc fragments. Antibody captured by the Fc assay is used as an antigen assay to determine the antibody-antigen binding stoichiometry. Intact IgG may bind a maximum of 2 antigens (orange circles), whereas samples containing fragments lacking Fab regions will show a lack of antigen capture. The sensor response of the antigen surface (orange) and PAG surface (blue) to the NISTmAb is shown, with and without the inclusion of the antigen assay step (solid and dotted lines respectively) to highlight the signal of antigen capture

SEC, sodium dodecyl sulphate-polyacrylamide gel electrophoresis (SDS-PAGE), and mass spectrometry (MS) are the gold standard techniques used for characterization and stability assessment of antibodies [4, 14, 15]. Non-reducing SDS-PAGE analysis of a papain digestion of the NISTmAb shows bands for Fab and Fc fragments at 50 kDa, whilst under reducing conditions, the disulphide bonds between heavy and light chains are separated to produce two species of 25 kDa [16]. The primary amino acid sequence and structure of antibodies are readily confirmed by liquid chromatography and mass spectrometry (LC-MS/MS) [17]; however, there are no easy methods to establish the monomeric purity or biological activity of a sample immediately before use. Therefore, in the current study, we consider the role of a multiplexed plasmonic biosensor-based assay for rapid monitoring of antibody monomeric purity in a research setting with similar analysis times to MS.

Plasmonic biosensors are typically used to observe antibody binding kinetics and are fundamentally mass sensors. Accurate kinetic parameters may be derived using antibody samples of high monomeric purity, careful tethering chemistry, and fitting of a mathematical kinetic model to multiple low-concentration interactions simultaneously in a global fit [18, 19]. However, kinetic analyses are fundamentally limited by unknown antibody concentration and monomeric purity. The presence of multiple binding interactions in fragmented antibody samples invalidates the assumed Langmuirian 1:1 binding interaction model which is commonly fit to plasmonic data [18].

In this paper, we report a novel multiplex technique to simultaneously assess antibody affinity and monomeric purity, based on our in-house biosensor platform [19, 20]. Sensor surfaces are functionalised with either antigen or the Fc binding protein A/G (PAG) to generate assays for Fab and Fc fragments. PAG has 3 sterically available binding sites for the Fc region of intact IgG and binds with high affinity to antibodies from several species [20]. The technique described here is comprised of two steps, shown in Fig. 1. In the first step, antibody binding to the Fab assay provides data for kinetic analysis of the antibody-antigen interaction and the mass of antibody binding to the Fc assay is recorded simultaneously. For a constant number of binding sites on the Fc assay surface, an initial estimate of monomeric purity may be obtained by observing a mass deficit in degraded samples compared to the NISTmAb. In the second step, the mass of antigen which can bind to the antibodies immobilized on the PAG surface (an antigen assay) is recorded. The stoichiometry of the antibody-antigen binding reaction on the PAG surface is derived from the masses of antibody and antigen, and samples with high monomeric purity should bind an average of 1–2 antigens per antibody, whereas antibody fragments lacking Fab regions should display a lower stoichiometry. A panel of eight antibody samples was tested: the NISTmAb, a therapeutic infliximab biosimilar, and six commercially available monoclonal and polyclonal samples for research use. Our results are validated with native ESI-quadrupole time-of-flight (QTOF) mass spectrometry—a gold standard technique for assessing the relative abundance of antibody fragmentation products in a sample [21].

Materials and methods

Pierce recombinant protein A/G (#21186) and anti-C5a C17/5 (#GAU025-05-02) were obtained from Thermo Scientific. Bovine serum albumin (BSA) (#A7030) was obtained from Sigma-Aldrich. NIST RM 8671 was purchased from the National Institute of Standards and Technology and respiratory syncytial virus protein F (RSVF, #11049-V08B) was obtained from Sino Biological Inc. C5a protein (#A145) was purchased from Complement Technologies Inc. Anti-CRP monoclonal, anti-CRP polyclonal, anti-TBG, TNFα, CRP, and TBG were all obtained from a major research reagent provider. Infliximab was obtained from clinical storage, reflecting the conditions of therapeutic antibodies before use. Standard instrument running buffer and sample dilution buffer was phosphate-buffered saline (PBS) supplied in tablet form by Sigma-Aldrich. Ortho-phosphoric acid (85%) was obtained from Fluka and a 0.01 M aqueous solution used as regeneration buffer. The absorbance of the NISTmAb and other antibody samples was measured at 280 nm with a NanoDrop instrument and converted to total protein concentration using Beer’s Law and a molar extinction coefficient of 210,000 M−1 cm−1 (as recommended by the manufacturer).

The binding kinetics and monomeric purity of 8 antibody samples was assessed using our novel biophotonic multiplexed biosensor platform, described in detail elsewhere [19, 20, 22]. Briefly, gold nanoparticles are printed onto an aminated glass surface using an inkjet printer and the nanoparticles are chemically grown on the surface to ~60 nm in diameter. The gold nanoparticles are illuminated in a total internal reflection configuration with the localized nanoparticle plasmon scattering light normal to the surface depending on the refractive index (mass) of protein in the plasmon field. The scattered light intensity is monitored in real time using a video camera capturing data at 30 frames per second.

A multiplexed biophotonic array is fabricated by functionalising each of the 150 array spots with proteins using EDC/NHS coupling chemistry [22]. The array contains some reference spots functionalised with BSA which are used to correct for variations in light intensity, temperature, and non-specific binding during the time course of the experiment. Spots functionalised with antigen are used for specific antibody capture via the Fab region (a Fab assay) and spots functionalised with PAG are used for antibody capture via the Fc region (an Fc assay). PAG binds to IgG in an orientation presenting the Fab regions away from the sensor surface into the solution, as seen in affinity column chromatography. Once loaded with a given antibody, the PAG surface may be used to capture antigen from a sample (an antigen assay).

A multiplexed array was printed with the Fc and Fab assays: PAG and antigen were printed with 23 repeats each from 1 mg/mL solutions in PBS. The control assay elements are printed with 25 repeats from a 1 mg/mL solution of BSA in PBS. The responses of assay elements were averaged to improve the signal-to-noise ratio and eliminate the effects of any inhomogeneous protein tethering density. The array surface was blocked with BSA (1 mg/mL, 300 s) before use to minimize non-specific binding. The response of each assay to refractive index (RI) change was calibrated with a step change in running buffer from PBS to 2 × PBS concentration with a known bulk RI change of 1.6 mRIU. The mean limit of detection per individual sensor element was 6.6 (± 2.9) × 10−2 mRIU.

The Fc assay is calibrated on each array by injecting the NISTmAb at concentrations of 1.56, 3.23, and 6.25 nM and the response of the Fc and Fab assays recorded simultaneously. The data were used for kinetic analysis of the NISTmAb with a global fit to the Langmuir adsorption isotherm [19, 20]. The arrays were regenerated with a 0.01 M solution of phosphoric acid for 100 s between samples. The PAG surface has been demonstrated previously to be stable over > 20 regeneration cycles [20]. For monomeric purity testing, all antibody samples were diluted to 100 nM (based on manufacturer stated concentration) and injected over the assay surface for a fixed period of 300 s to monitor association kinetics to the Fab assay and Fc assay. The assays were washed in PBS running buffer for 500 s to monitor dissociation kinetics. Each antigen sample was prepared at 100 nM in PBS and injected over the assay surfaces for 300 s to observe antigen capture by IgG previously loaded on the PAG surface.

The sensor response to antibody or antigen binding at maximum surface coverage, ϑm, was derived for all experiments with kinetic fitting of the Langmuir adsorption isotherm. All parameters estimated from kinetic fitting are presented with their associated 95% confidence interval derived from the covariance matrix of the fit. For clarity, the ϑmvalue of the Fab assay is written ϑm Fab and the ϑmvalue of the Fc assay similarly ϑm Fc. The ϑmvalue of the antigen assay is termed ϑm antigen. To minimize experimental variation, all antibody samples were tested in duplicate and in random sequence on a single multiplex array of assay surfaces. The coefficient of variation (CV) of ϑm Fc for the NISTmAb was 7%.

The LC-MS analysis was conducted using an HPLC-Chip Cube system coupled to a 6520 quadrupole time-of-flight (QTOF) mass spectrometer (Agilent Technologies, Santa Clara, CA) operated in electrospray ionization (ESI) positive-ion mode. Liquid chromatography was performed using a Protein-Chip (II) with a 40 nL enrichment column and analytical column of 43 mm × 75 μM with Zorbax 300SB-C8 packing material at 5 μm (G4240-63001, Agilent, Santa Clara, CA). The ChipCube source was operated at 365 °C with 5 L/min nitrogen drying gas, the capillary voltage was set to 2100 V and fragmentor at 400 V. The TOF MS acquisition range was from 100 to 4000 mass-to-charge ratio (m/z) at 13586 transients per summed spectra. The source was interfaced with an Agilent 1260/1200 series HPLC system consisting of a 1260 Cap pump, 1200 Nano pump, 1200 Micro WPS, and 1290 Infinity Thermostat (Agilent). Between 0.2 and 1 μL sample, without dilution or additional preparation, was injected onto the enrichment column using the capillary pump flow with H2O + 0.1% formic acid (FA) at a flow rate of 4 μL/min. The sample was eluted onto the analytical column at a flow rate of 0.6 μL/min. Solvent A and B consisted of H2O + 0.1% formic acid (FA) and ACN:H2O 90:10 with 0.1% FA respectively. Gradient steps are as follows: 0–4 min from 3% B to 50% B, 4–5 min to 100% B, 5–11 min 100% B, 11–12 min from 100% to 3% B. Internal lock mass calibration was performed using one calibration reference mass at 1221.9906 m/z. Data processing was performed using the Masshunter Workstation software version B.50.00 (Agilent). The mass spectra of each antibody sample are averaged from triplicate injections derived from two independent experiments.

The ESI-QTOF experiments analysed native, intact forms of the antibodies. The raw MS data undergoes a zero-charge convolution in which the area contribution of each respective charge state is summed [23]. We expressed the degree of antibody degradation as a relative area ratio of the light or heavy chain combinations compared to intact protein in the deconvoluted spectra. As we are relating relative area ratios of the same antibody in each sample, the ionization efficiency and therefore relative abundance estimates are expected to be internally consistent.

Results

A series of kinetic traces was recorded for each antibody sample for the Fc, Fab, and antibody-antigen assays. The data are fitted to a Langmuir model with integrated rate equations given by Eq. 1 for the association phase and Eq. 2 for the dissociation phase:
$$ \upvartheta (t)={\upvartheta}_m\frac{k_a\left[P\right]}{k_a\left[P\right]+{k}_d}\left(1-{e}^{-\left({k}_a\left[P\right]+{k}_d\right)t}\right) $$
(1)
$$ \upvartheta (t)={\upvartheta}_a{e}^{-{k}_dt} $$
(2)
where ϑ(t) is the sensor response at time, t, ϑm as before, ϑa is the sensor response at the start of the dissociation phase, ka is the association rate constant, kd is the dissociation rate constant, and [P] is the concentration of the protein in solution whether Fc, Fab, or antigen. All of the parameters in Eqs. 1 and 2 are highly correlated when fitted to the data of a single concentration. However, the binding rate constants ka and kd may be determined by fitting the model to multiple, low-concentration samples simultaneously [19]. Furthermore, the ϑm can be determined from replicates of a high concentration sample (100 nM) binding over a time sufficient to achieve near-maximum surface coverage. Simultaneous low-concertation kinetic trace fitting and independent determination of ϑm give the most accurate determination of the interaction parameters in the Langmuir model.
The ϑm of the NISTmAb sample binding via the Fab region to its antigen (RSVF) immobilized on the sensor surface was derived as 1.73 (± 0.03) mRIU by fitting the Langmuir model to data from 100 nM antibody samples. The kinetic parameters for the NISTmAb–RSVF interaction were derived as ka = (8.9 ± 0.2) × 104 M−1 s−1, kd = (2.4 ± 0.6) × 10−4 s−1, and KD = 2.7 ± 0.7 nM by fitting the Langmuir model to 3 low-concentration samples (Fig. 2). The uncertainties are derived from the covariance matrix of the fit and are stated as 95% confidence intervals.
Fig. 2

RI time dependence for the NISTmAb binding to its antigen, RVSF, immobilized on the sensor surface: a kinetic response curves obtained from 1.56, 3.23, and 6.25 (± 5%) nM NISTmAb samples (grey), and global fit to the Langmuir adsorption model (red); b residuals over the time course of the fit, showing consistent error between model fit and data; c histogram of the residuals indicating the error between the model fit and data is normally distributed.

The ϑm of the NISTmAb sample binding to PAG on the sensor surface was derived as 6.7 ± 1.0 mRIU by fitting the Langmuir model to data from 100 nM antibody samples. Deviations from 1:1 Langmuir kinetics were evident in the data however a global fit to four low concentrations, whilst enabling the ϑm value to float, enables the model to fit well and estimates the kinetics of the interaction as ka = (5.9 ± 1.1) × 104 M−1 s−1 and kd = (2.0 ± 1.9) × 10−5 s−1. The error associated with the dissociation rate constant is large due to the apparently slow dissociation rate.

The ϑm Fab, ϑm Fc, and ϑm antigen of all antibody samples and their analytes binding to the Fab, Fc, and antigen assays were estimated from binding data (Fig. 3) using the Langmuir fitting routine and the values are listed in Table 1. All ϑestimates assume antibody concentrations stated in the product data sheets which were confirmed as accurate by absorbance at A280 nm. The ϑm Fab is consistently lower than the ϑm Fc for all antibody samples tested, indicating that for a given surface area, more antibodies are adsorbed when bound via the Fc region to the PAG surface compared to the antigen surface. The therapeutic infliximab biosimilar antibody has a particularly low ϑm Fab value indicating a low antigen binding density or a low epitope presentation on the sensor surface.
Fig. 3

Biosensor analysis of antibody panel samples: ah raw kinetic response of the PAG surface (blue) and antigen surface (orange) to antibody binding (0–300 s) and antigen binding (400–700 s). i ϑm analysis showing the ϑm Fabm Fc ratio (green) and the ϑm antigenm Fc ratio (gold). The ϑm values of the ϑm antigenm Fc ratio are normalized for the mass of the antigen and intact antibody derived from the ESI data. Error bars are 95% confidence intervals. The stoichiometric limit of the normalized ϑm antigenm Fc ratio is between 1 and 2 (dashed lines).

Table 1

Experimental properties of the antibody panel. All errors are 95% confidence intervals

Sample

ϑm Fab/mRIU

ϑm Fc/mRIU

ϑm antigen/mRIU

Antigen mass/kDa

ESI mass, MESI/kDa

ESI monomeric purity/%

Plasmonic Fc areal density, Madsorbed/g cm−2 × 10−8

Plasmonic Fc mass,  MFc/kDa

Plasmonic monomeric purity/%

NISTmAB

1.73 ± 0.03

6.66 ± 0.96

3.37 ± 0.77

53

145

94

3.66 ± 0.53

148 ± 21

100 ± 14

Infliximab biosimilar

0.22 ± 0.06

6.82 ± 0.21

1.32 ± 0.01

17

146

95

3.8 ± 0.1

152 ± 5

103 ± 3

Anti-C5a mono

0.42 ± 0.13

3.07 ± 0.28

0.11 ± 0.04

10.4

142

89

1.7 ± 0.2

68 ± 6

46 ± 4

Anti-CRP mono

0.02 ± 0.02

3.57 ± 0.13

1.44 ± 0.68

125

141

87

1.96 ± 0.07

80 ± 3

54 ± 2

Anti-TNFα poly

2.34 ± 0.13

3.23 ± 0.88

0.42 ± 0.06

17

123

61

1.8 ± 0.4

72 ± 19

49 ± 13

Anti-C5a poly

5.82 ± 1.48

5.36 ± 0.60

0.06 ± 0.04

10.4

127

65

3.0 ± 0.3

119 ± 12

80 ± 8

Anti-CRP poly

0.76 ± 0.30

4.33 ± 0.09

1.18 ± 0.45

125

132

72

2.38 ± 0.05

96 ± 2

65 ± 1

Anti-TBG poly

1.60 ± 0.70

3.74 ± 0.24

0.08 ± 0.05

54

119

54

2.1 ± 0.1

83 ± 5

56 ± 3

An estimate of the relative binding site density differences between PAG and the antigens may be obtained by comparing the ϑm Fabm Fc ratio of the NISTmAb (0.26 ± 0.04) to that of the panel of antibodies with a range = (0.004–1.1) × 10−3. Furthermore, the ϑm antigenm Fc ratio of the antibody samples is an estimate of antibody monomeric purity (related to moles/moles), after normalizing for molecular mass by dividing ϑm antigen and ϑm Fc by the antigen masses and intact IgG masses respectively (Fig. 3i).

To validate the ϑm antigenm Fc ratio as a measure of antibody monomeric purity, the biosensor results were compared with results from the ESI mass spectrum recorded for each antibody (see Electronic Supplementary Material (ESM) Fig. S1). The deconvolved mass spectrum of the NISTmAb is dominated by a species at 148.2 kDa attributed to the pure, non-fragmented IgG structure. The NISTmAb Fab fragment has a reported mass of 47 (± 5) kDa [16] and the mass spectrum recorded here shows a low intensity mass peak at 48 kDa. All antibody samples show additional lower mass proteins with varying abundances attributed to any of the fragments shown in Fig. 1, including dissociated heavy chain (Hc) and light chain (Lc). The relative abundances of fragments relative to the intact antibody (Fig. 4) are determined from the area-under-the-curve (AUC) of regions in which masses are typically present: 140–160 kDa, 90–110 kDa, 65–85 kDa, 40–60 kDa, and 15–35 kDa. These regions of the mass spectrum correspond to the predicted fragmentation products shown in Fig. 1.
Fig. 4

The abundances of fragments in each antibody sample evaluated by taking the AUC of specific regions of the mass spectra corresponding to IgG fragmentation products. The percentage abundances of the most abundant peaks are plotted relative to the AUC of the intact region (140–160 kDa): 90–110 kDa (light blue), 65–85 kDa (gold), 40–60 kDa (pink), and 15–35 kDa (blue). All regions may contain Fc moieties.

The HcLc fragment, expected at 75 kDa, is the most abundant fragmentation product in all monoclonal antibody samples, whilst the Fab-Fc fragment is most abundant in the polyclonal samples, with the exception of anti-TBG. The AUC of impurities relative to the intact IgG AUC is < 20% for monoclonal antibodies. In contrast, all polyclonal samples display impurities present at > 20% relative to the intact IgG structure. The monomeric purity estimated by ESI is the peak area of the intact region expressed as a percentage of total peak area (area/area) and is tabulated in Table 1, expressed as a mean of 3 mass spectra and a 95% confidence limit of ± 2%. The monomeric purity of the NISTmAb is determined by ESI to be 94 ± 2% and compares well to the reported value 97% by NIST [8].

Discussion

Antibody monomeric purity is critical for immunoassay development, binding kinetics studies, and effective immunotherapy. The Fab assay provides data for kinetic analysis of the antibody-antigen binding reaction, whilst the Fc and antigen assays provide potential assays for the rapid assessment of antibody monomeric purity. The Fc assay may identify low monomeric purity based on a potential mass deficit observed for fragmented samples binding to a constant number of binding sites on the PAG surface. The antigen assay assesses monomeric purity using the binding ratio of antigen to antibody captured by the PAG surface. Results are validated using monomeric purity estimates from native ESI-QTOF mass spectrometry.

ESI data suggest the most abundant glycoform of the intact NISTmAb has a mean molecular mass of 148.2 kDa and the sample displays a distinctive glycosylation pattern in good agreement with published values [11]. The homogeneity of the NISTmAb contrasts with the heterogeneity of intact IgG masses in the anti-C5a monoclonal, anti-CRP monoclonal, and anti-CRP polyclonal samples, attributed to multiple glycoforms (ESM Fig. S2) [24]. The ESI data validate the NISTmAb as a good standard antibody of high monomeric purity. From the ESI mass spectrum data, the mean mass of proteins in the antibody samples, MESI, can be derived:
$$ {M}_{ESI}=\frac{aA+ bB+ cC+ dD}{a+b+\mathrm{c}+\mathrm{d}} $$
(3)
where lowercase letters represent the percentage AUC of the mass spectrum relative to the intact peak area (140–160 kDa) from the regions: 90–110 kDa (predicted Fab-Fc complex), 65–85 kDa (predicted HcLc complex), 40–60 kDa (predicted Hc or Lc), and 15–35 kDa (predicted Lc or Hc monomer of Fc). Uppercase letters represent the mean mass of the integration region. The derived MESI values of each sample in the panel are shown in Table 1.

The monomeric purity of the NISTmAb and the therapeutic infliximab biosimilar is high, which is not the case for the other commercially available samples that would be used for research applications. The reliability of experiments involving Western blots, immunohistochemistry, flow cytometry, and immunoprecipitation is negatively impacted by variations in antibody purity as these techniques fundamentally rely on a connection between Fc and Fab regions of the antibody. The consequence of purity on kinetic analysis also needs to be considered with the fundamental errors of the 1:1 binding model or Langmuirian kinetics conventionally used to analyse antibody binding events. Fitting a set of low concentrations in the global fit methodology typically produces an error of 15% for ka and 10% for kd over the timescale of the experiments [19]. Model fitting errors would dominate the analysis of the NISTmAb and infliximab biosimilar samples but this would not be true for all members of the antibody panel tested here. For the case of the anti-TBG sample, any reported antibody kinetics analyses without a sample purity assessment and correction may have an error of up to 45% due to low monomeric purity.

The packing density of IgG around a protein is a property of epitope density, antibody conformation, and monomeric purity. Antibodies with identical masses should achieve identical ϑm values for a given antibody binding density, albeit after differing periods of time to allow variations in association kinetics. There is significant variation in the ϑm Fab values recorded (Table 1) which may be explained by variation in availability of epitopes for antibody binding as evidenced by the larger ϑm Fab of polyclonal anti-CRP compared to the monoclonal anti-CRP sample on the same CRP surface. In contrast, the ϑm Fc shows less variation than ϑm Fab, attributed to the use of a common PAG surface for all antibody samples with a consistent number of binding sites.

Variation in ϑm Fc may occur due to differing mean IgG mass differences (monomeric purity differences) between IgG samples as observed with ESI. The Fc assay presented here will bind with high affinity to both intact IgG and the most abundant fragmentation products which contain the Fc region (Fig. 1), a concern for PAG-based affinity purification of antibodies but a potential antibody fragmentation assay when combined with surface plasmon technology. Any given ϑm derived from fitting of the Langmuir equation (Eqs. 1 and 2) may be related to the adsorbed mass on the surface, Madsorbed, by de Feijter’s formula [25]:
$$ {M}_{\mathrm{adsorbed}}={d}_A\frac{\Delta RI}{\raisebox{1ex}{$ dn$}\!\left/ \!\raisebox{-1ex}{$ dc$}\right.} $$
(4)
where dA is the thickness of the adsorbed layer, dn/dc is the rate of change of refractive index with surface concentration (the refractive index increment), and ΔRI is the change in the refractive index recorded. The value of dn/dc has been measured for protein as 0.182 g cm−3 and has been shown to vary with buffer conditions [26] but the effect is small [25].

The largest uncertainty in the estimate of Madsorbed is the value of dA. IgG layers adsorbed onto protein A and silica surfaces have been observed previously using ellipsometry with thicknesses of 4 nm [27] to 16 nm [28] respectively. Neutron reflectivity measurements of IgG4 binding to a protein A/BSA surface report three sequential layers leading to total thicknesses between 6.1 nm and 25 nm depending on washing and blocking of the surface [29] because non-specific 3D aggregation at high IgG concentrations can enable adsorbed layers to exceed the maximum length of a single IgG molecule [30]. The mean cross-section dimension of the IgG crystal structure [31] suggests a dA of 10 nm, assumed constant for all antibody samples here. The derived surface adsorbed mass density therefore depends only on the difference in the refractive index (∆RI) of the adsorbed antibody.

Substituting ϑm Fc of NISTmAb for ∆RI in Eq. 4 predicts the areal density of the NISTmAb on the PAG surface, Madsorbed = 3.7 ± 0.5 × 10−8 g cm−2. By comparison, electrostatic adsorption of monoclonal antibodies onto a variety of surfaces has been reported at densities in the range 5 × 10−7 to 5 × 10−6 g cm−2 [32, 33]. Electrostatic adsorption is expected to achieve a higher surface mass density compared with the ordered two-dimensional layer of IgG bound to a PAG surface. The Madsorbed values of each antibody studied are presented in Table 1.

The ESI data indicate a high monomeric purity in the NISTmAb sample (94% ± 2) with an intact mass, Mmonomer, of 148.2 kDa. The NISTmAb data and the Avogadro constant, NA, can be used to calibrate the surface density of antibodies bound at saturation, NAb = Madsorbed × NA/Mmonomer = (1.49 ± 0.21) × 1011 cm−2. Comparatively, the maximum number of monoclonal antibodies in a monolayer has been determined by adsorption onto a silicon nitride surface as 1.2 × 1012 cm−2 [34]. The unknown purity of antibody samples in the literature is a potentially large source of error for published NAb values derived from protein mass density measurements. The NAb figure presented here is of known accuracy due to the monomeric purity estimate from ESI.

The number of PAG binding sites on the surface is constant for NISTmAb and if all antibodies bind to the surface with the same packing density then an estimate of effective mass at each binding site may be estimated from a simple ratio:
$$ {\upvartheta}_m^{Fc}={\upvartheta}_m^{\mathrm{monomer}}\times \frac{M_{Fc}}{M_{\mathrm{monomer}}} $$
(5)
where \( {\upvartheta}_m^{\mathrm{monomer}} \) and Mmonomer are the ϑm Fc and mass of the monomeric reference material, which in the present study is the NISTmAb with ϑm Fc = 6.66 ± 0.96 mRIU and a mass of 148.2 kDa. When an antibody sample has degraded, a fraction of Fc binding domains on the PAG will be occupied by antibody fragments with a lower mean molecular mass than the intact IgG reference and will exhibit a lower ϑm Fc for a constant number of occupied binding sites. The Fc assay can be used to estimate monomeric purity by expressing MFc as a percentage of Mmonomer (Table 1).

A weak positive correlation is observed between ϑm Fc and ESI monomeric purity and several phenomena may account for the variations observed in ϑm Fc. Firstly, NAb at ϑm Fc may be a fundamental property of a given antibody due to glycan variation and the locations of PAG binding domains on the structure, leading to some antibody monomers binding in orientations which favour closer packing on the surface [35]. The ϑm Fc will also be sensitive to non-specific aggregation and clustering of IgG [29], although aggregates of the NISTmAb IgG are only observed following deliberate physical agitation for days at room temperature [36] and the antibody samples tested here are not expected to contain significant aggregates due to appropriate storage. The lack of correlation between ϑm Fc and ESI monomeric purity may best be explained by the orientation and higher packing density of low mass antibody degradation products, which may have access to a greater number of binding sites on the PAG surface than monomeric IgG [20]. Therefore, it is likely that the ϑm Fc of the Fc assay is only sensitive to severe antibody degradation.

Our results show that a more sensitive measure of monomeric purity is to support antibodies on the PAG surface and measure the antigen binding ratio, ϑm antigenm Fc, which after normalizing for molecular mass, shows a strong dependence on monomeric purity estimated by ESI (Fig. 5).
Fig. 5

The relationship between monomeric purity estimated from the ESI data and the mass -normalized ϑm antigenm Fc biosensor measurement. The NISTmAb and infliximab biosimilar (green) are > 90% pure according to ESI data and exhibit ϑm antigenm Fc ratios within the expected limits of between 1 and 2 antigens per antibody. Other commercially available antibodies (red) are < 90% pure and show less than 1 antigen bound per antibody. A 5-parameter logistic curve fits well to the data (blue) and is constrained by a maximum asymptote of 2 (the upper limit of the antigen-antibody interaction stoichiometry). The anti-TNFα polyclonal (orange) does not fit the trend line, likely due to a unique fragmentation pattern resulting in a majority of 100 kDa HcLc fragments which may pack tightly on the PAG surface and remain capable of binding 1 antigen

The infliximab biosimilar antibody binds to a relatively low mass antigen (TNFα with a monomer mass of 17 kDa) and displays an antigen-antibody binding ratio of 1.71 ± 0.16. The NISTmAb binds to a higher-mass antigen (53 kDa) and displays a lower antigen-antibody binding ratio of 1.31 ± 0.57 attributed to greater steric hindrance of antigens. The ϑm antigen/ϑm Fc ratio threshold to determine monomeric purity of antibody samples is antigen-dependent for larger antigens but would always exceed 1 with adequate antibody spacing on a plasmonic surface. The NISTmAb performs well as a standard material for the antibodies shown, but calibration of the assay would likely be improved by comparing each tested sample to a high-purity reference standard which binds to a common antigen.

Of the commercially available research antibodies, anti-TBG antibody shows a relatively high ϑm Fab/ϑm Fc ratio but displays the weakest ϑm antigenm Fc ratio. These two ratios indicate that both Fab and Fc regions of TBG are present in the sample, but a significant proportion is separated due to degradation. The high fragmentation of the anti-TBG antibody is supported by the lack of intact antibody in the ESI mass spectrum (ESM Fig. S1).

The ϑm antigenm Fc ratio is a measure of the biological activity of an antibody sample, whereas the ESI monomeric purity estimated from mean mass penalizes any form of antibody fragmentation. The ESI data predict a low monomeric purity in the anti-TNFα polyclonal but the sample displays a relatively high ϑm antigenm Fc ratio. The result may be explained by the unique degradation pattern shown in Fig. 4 which indicates that the most abundant species is the 100 kDa HcLc fragment which is capable of binding to PAG and one antigen. The fragmentation in the anti-TNFα polyclonal remains identifiable by comparison to the infliximab biosimilar which shares the same antigen but exhibits a higher antigen-antibody binding ratio.

Using the ϑm Fab/ϑm Fc ratio recorded by the biosensor, the antibody panel readily divides into two types of samples: the high-purity (> 90%) samples (NISTmAb and infliximab biosimilar) and the remainder of the panel with lower monomeric purity (< 90%). The ϑm Fab/ϑm Fc ratio is significantly different for these two material types as evidenced by a two-sample t test and two-sample Kolmogorov-Smirnov (P = 0.01 and P = 0.03 respectively). The assays developed here show the potential to test the specification of antibodies immediately before use as assay materials and in other antibody-dependent techniques to prevent variability in experimental results caused by sample degradation. Both the ESI and biosensor-based assays provide an estimate of monomeric purity in 15 min but the biosensor platform exhibits some key advantages. The biological activity of therapeutic antibody at both the Fab and Fc regions is critical and can be tested on the small form factor multiplexed biosensor platform when reliably calibrated using the reference NISTmAb.

Conclusions

The NISTmAb provides an important new benchmark material which enables the calibration of biosensors assessing monomeric purity of antibody samples with assays for Fab, Fc, and antigen. The analysis of a panel of commercially available research grade antibodies, using ESI and biosensor surfaces, highlights significant variation in monomeric purity and raises accuracy concerns for previously reported ka, kd, and KD measurements of similar materials in the literature and the risk to reproducibility. Without sample monomeric purity analysis (of both the antibody and the antigen) these measurements may have errors approaching 50% in highly fragmented samples. The biosensor-based technique presented here to estimate antibody monomeric purity can be performed in 15 min and may be combined with additional rapid tests for IgG aggregates as described for the NISTmAb in ref. [36].

Notes

Acknowledgements

PP Reader would like to thank the EPSRC for a PhD Studentship and Attomarker Ltd. for access to equipment during the PhD.

Compliance with ethical standards

Conflict of interest

Prof. Shaw is a Director of Attomarker Ltd that owns the core IP for the array reader technology. No other authors have any conflicts of interests to declare.

Supplementary material

216_2019_2029_MOESM1_ESM.pdf (428 kb)
ESM 1 (PDF 428 kb)

References

  1. 1.
    Baker M. Reproducibility crisis: blame it on the antibodies. Nature. 2015;521(7552):274–6.CrossRefPubMedGoogle Scholar
  2. 2.
    Weller MG. Quality issues of research antibodies. Anal Chem Insights. 2016;11:21–7.CrossRefPubMedPubMedCentralGoogle Scholar
  3. 3.
    Baker M. Antibody anarchy: a call to order. Nature. 2015;527:545.CrossRefPubMedGoogle Scholar
  4. 4.
    Vlasak J, Ionescu R. Fragmentation of monoclonal antibodies. mAbs. 2011;3(3):253–63.CrossRefPubMedPubMedCentralGoogle Scholar
  5. 5.
    Correia IR. Stability of IgG isotypes in serum. mAbs. 2010;2(3):221–32.CrossRefPubMedPubMedCentralGoogle Scholar
  6. 6.
    Voskuil JLA. The challenges with the validation of research antibodies. F1000Research. 2017;6:161.CrossRefPubMedPubMedCentralGoogle Scholar
  7. 7.
    Points to consider in the manufacture and testing of monoclonal antibody products for human use (1997). U.S. Food and Drug Administration Center for Biologics Evaluation and Research. J Immunother. 1997;20(3):214–43.Google Scholar
  8. 8.
    Schiel JE, Turner A, Mouchahoir T, Yandrofski K, Telikepalli S, King J, et al. The NISTmAb reference material 8671 value assignment, homogeneity, and stability. Anal Bioanal Chem. 2018;410(8):2127–39.CrossRefPubMedPubMedCentralGoogle Scholar
  9. 9.
    Turner A, Yandrofski K, Telikepalli S, King J, Heckert A, Filliben J, et al. Development of orthogonal NISTmAb size heterogeneity control methods. Anal Bioanal Chem. 2018;410(8):2095–110.CrossRefPubMedPubMedCentralGoogle Scholar
  10. 10.
    Schiel JE, Mire-Sluis A, Davis D. Monoclonal antibody therapeutics: the need for biopharmaceutical reference materials. State-of-the-Art and Emerging Technologies for Therapeutic Monoclonal Antibody Characterization Volume 1 Monoclonal Antibody Therapeutics: Structure, Function, and Regulatory Space. ACS Symposium Series. 1176: Am Chem Soc; 2014. p. 1–34.Google Scholar
  11. 11.
    Formolo T, Ly M, Levy M, Kilpatrick L, Lute S, Phinney K, et al. Determination of the NISTmAb primary structure. State-of-the-Art and Emerging Technologies for Therapeutic Monoclonal Antibody Characterization Volume 2 Biopharmaceutical Characterization: The NISTmAb Case Study. ACS Symposium Series. 1201: Am Chem Soc; 2015. p. 1–62.Google Scholar
  12. 12.
    McLellan JS, Chen M, Kim A, Yang Y, Graham BS, Kwong PD. Structural basis of respiratory syncytial virus neutralization by motavizumab. Nat Struct Mol Biol. 2010;17(2):248–50.CrossRefPubMedPubMedCentralGoogle Scholar
  13. 13.
    Cohen SL, Price C, Vlasak J. β-Elimination and peptide bond hydrolysis: two distinct mechanisms of human IgG1 hinge fragmentation upon storage. J Am Chem Soc. 2007;129(22):6976–7.CrossRefPubMedGoogle Scholar
  14. 14.
    Liu H, Gaza-Bulseco G, Sun J. Characterization of the stability of a fully human monoclonal IgG after prolonged incubation at elevated temperature. J Chromatogr B Anal Technol Biomed Life Sci. 2006;837(1–2):35–43.CrossRefGoogle Scholar
  15. 15.
    Zhang H, Cui W, Gross ML. Mass spectrometry for the biophysical characterization of therapeutic monoclonal antibodies. FEBS Lett. 2014;588(2):308–17.CrossRefPubMedGoogle Scholar
  16. 16.
    Karageorgos I, Gallagher ES, Galvin C, Gallagher DT, Hudgens JW. Biophysical characterization and structure of the Fab fragment from the NIST reference antibody, RM 8671. Biologicals. 2017;50:27–34.CrossRefPubMedPubMedCentralGoogle Scholar
  17. 17.
    Resemann A, Jabs W, Wiechmann A, Wagner E, Colas O, Evers W, et al. Full validation of therapeutic antibody sequences by middle-up mass measurements and middle-down protein sequencing. mAbs. 2016;8(2):318–30.CrossRefPubMedPubMedCentralGoogle Scholar
  18. 18.
    Latour RA. The langmuir isotherm: a commonly applied but misleading approach for the analysis of protein adsorption behavior. J Biomed Mater Res A. 2015;103(3):949–58.CrossRefPubMedGoogle Scholar
  19. 19.
    Read T, Olkhov RV, Williamson ED, Shaw AM. Label-free Fab and Fc affinity/avidity profiling of the antibody complex half-life for polyclonal and monoclonal efficacy screening. Anal Bioanal Chem. 2015;407(24):7349–57.CrossRefPubMedGoogle Scholar
  20. 20.
    Reader PP, Shaw AM. Kinetic analysis of the multivalent ligand binding interaction between protein A/G and IgG: a standard system setting. J Phys Chem B. 2017;121(38):8919–25.CrossRefPubMedGoogle Scholar
  21. 21.
    Beck A, Sanglier-Cianférani S, Van Dorsselaer A. Biosimilar, biobetter, and next generation antibody characterization by mass spectrometry. Anal Chem. 2012;84(11):4637–46.CrossRefPubMedGoogle Scholar
  22. 22.
    Olkhov RV, Fowke JD, Shaw AM. Whole serum BSA antibody screening using a label-free biophotonic nanoparticle array. Anal Biochem. 2009;385(2):234–41.CrossRefPubMedGoogle Scholar
  23. 23.
    Rosati S, Yang Y, Barendregt A, Heck AJR. Detailed mass analysis of structural heterogeneity in monoclonal antibodies using native mass spectrometry. Nat Protoc. 2014;9:967.CrossRefPubMedGoogle Scholar
  24. 24.
    Higel F, Seidl A, Sorgel F, Friess W. N-glycosylation heterogeneity and the influence on structure, function and pharmacokinetics of monoclonal antibodies and Fc fusion proteins. Eur J Pharm Biopharm. 2016;100:94–100.CrossRefPubMedGoogle Scholar
  25. 25.
    Vörös J. The density and refractive index of adsorbing protein layers. Biophys J. 2004;87(1):553–61.CrossRefPubMedPubMedCentralGoogle Scholar
  26. 26.
    Ball V, Ramsden JJ. Buffer dependence of refractive index increments of protein solutions. Biopolymers. 1998;46(7):489–92.CrossRefGoogle Scholar
  27. 27.
    Wang D-S, Chang C-C, Shih S-C, Lin C-W. An ellipsometric study on the density and functionality of antibody layers immobilized by a randomly covalent method and a protein A-oriented method. Biomed Eng-App Bas C. 2009;303–10 p.Google Scholar
  28. 28.
    Lassen B, Malmsten M. Structure of protein layers during competitive adsorption. J Colloid Interface Sci. 1996;180(2):339–49.CrossRefGoogle Scholar
  29. 29.
    Mazzer AR, Clifton LA, Perevozchikova T, Butler PD, Roberts CJ, Bracewell DG. Neutron reflectivity measurement of protein A–antibody complex at the solid-liquid interface. J Chromatogr. 2017;1499:118–31.CrossRefGoogle Scholar
  30. 30.
    Clark NJ, Zhang H, Krueger S, Lee HJ, Ketchem RR, Kerwin B, et al. Small-angle neutron scattering study of a monoclonal antibody using free-energy constraints. J Phys Chem B. 2013;117(45):14029–38.CrossRefPubMedGoogle Scholar
  31. 31.
    Tan YH, Liu M, Nolting B, Go JG, Gervay-Hague J, Liu G-y. A nanoengineering approach for investigation and regulation of protein immobilization. ACS Nano 2008;2(11):2374–2384.Google Scholar
  32. 32.
    Stenberg E, Persson B, Roos H, Urbaniczky C. Quantitative determination of surface concentration of protein with surface plasmon resonance using radiolabeled proteins. J Colloid Interface Sci. 1991;143(2):513–26.CrossRefGoogle Scholar
  33. 33.
    Schlatter D, Barner R, Fattinger C, Huber W, Hübscher J, Hurst J, et al. The difference interferometer: application as a direct affinity sensor. Biosens Bioelectron. 1993;8(2):109–16.CrossRefGoogle Scholar
  34. 34.
    Williams RA, Blanch HW. Covalent immobilization of protein monolayers for biosensor applications. Biosens Bioelectron. 1994;9(2):159–67.CrossRefPubMedGoogle Scholar
  35. 35.
    Hanson DC, Schumaker VN. A model for the formation and interconversion of protein A-immunoglobulin G soluble complexes. J Immunol. 1984;132(3):1397.PubMedGoogle Scholar
  36. 36.
    Cheung CSF, Anderson KW, Patel PM, Cade KL, Phinney KW, Turko IV. A new approach to quantification of mAb aggregates using peptide affinity probes. Sci Rep. 2017;7:42497.CrossRefPubMedPubMedCentralGoogle Scholar

Copyright information

© The Author(s) 2019

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Authors and Affiliations

  • Peter P. Reader
    • 1
    • 2
  • Rouslan V. Olkhov
    • 2
  • Shaun Reeksting
    • 3