Analytical and Bioanalytical Chemistry

, Volume 410, Issue 22, pp 5507–5519 | Cite as

Pesticide analysis by pulsed flow modulation GCxGC-MS with Cold EI—an alternative to GC-MS-MS

  • Uri Keshet
  • Paulina Goldshlag
  • Aviv Amirav
Research Paper
Part of the following topical collections:
  1. Food Safety Analysis

Abstract

We explored the use of pulsed flow modulation (PFM) two-dimensional comprehensive gas chromatography (GCxGC) mass spectrometry with supersonic molecular beams (SMB) (also named Cold electron ionization (EI)) for achieving universal pesticide analysis in agricultural products. The use of GCxGC serves as an alternative to MS-MS in the needed reduction of matrix interference while enabling full-scan MS operation for universal pesticide analysis with reduced number of false negatives. Matrix interference is further reduced with Cold EI in view of the enhancement of the molecular ions. Pulsed flow modulation is a simple GCxGC modulator that does not consume cryogenic gases while providing tuneable second GCxGC column injection time for enabling the use of quadrupole-based mass spectrometry regardless its limited scanning speed. PFM-GCxGC-MS with Cold EI combines improved separation of GCxGC with Cold EI benefits of tailing-free ultra-fast ion source response time and enhanced molecular ions for the provision of increased sample identification information and reduced matrix interference. Consequently, PFM GCxGC-MS with Cold EI also improved NIST library identification probabilities of the spiked pesticides. PFM GCxGC is further characterized by largely increased second column sample and matrix capacity that as a result performs much better than thermal modulation GCxGC-MS with standard EI in the suppression of matrix interference. In a comparison with standard GC-MS, we measured with PFM GCxGC-MS with Cold EI an average total ion count matrix interference reduction factor of 32 for 12 pesticides in two matrices of baby leaves mixture and lettuce. In addition, Cold EI further increases the range of pesticides amenable for GC-MS analysis and its response is relatively uniform hence with it the need for pesticides specific calibration is reduced.

Graphical abstract

Pulsed flow modulation GCxGC-MS with Cold EI significantly reduces matrix interference and improves sample identification.

Keywords

Pulsed flow modulation GCxGC GCxGC-MS Supersonic molecular beams Cold EI Pesticide analysis 

Introduction

Pesticide analysis in agricultural products is an application area of growing importance which is usually addressed by GC-MS and LC-MS. In the British Crop Production Council’s (BCPC) printed Pesticides Manual, an estimated 1556 chemicals are being used as pesticides worldwide, including banned pesticides and other hazardous chemicals [1]. In the online database, there are above 10,400 names of pesticides, from which about 3100 are discontinued or banned [2]. In view of international food trade, this large number implies that pesticide analysis should not be treated as a limited number of target compounds analysis. Furthermore, pesticide analysis is performed in complex agricultural product matrices that severely impede the achievable limits of detection via extended matrix interference. As a result, the use of MS-MS is growing owing to its reduction of matrix interference and GC-MS-MS is now offered by all the major GC-MS vendors. While we did not find any paper that discusses the subject of by how much on the average (average for all relevant pesticides and matrices) MS-MS reduces the matrix interference, we estimate the MS-MS matrix interference reduction factor to be about 20 times in GC-MS versus the four ions reconstructed single ion monitoring (RSIM) method and it lowers the limit of detection (LOD) by about an order of magnitude. MS-MS is by definition a target analysis method and as a result, it is limited to only few hundred pre-selected (target) pesticides. Consequently, MS-MS the same as SIM methods are rigid and are guaranteed to miss any non-targeted pesticide residues that are not included in its list (most pesticides) thereby inherently leading to false negatives.

Thus, there is a growing need for a new method of pesticide analysis that will combine the MS-MS strength of matrix interference reduction with the features of full scan for the analysis capability of unlimited number of GC-MS amenable pesticides. As a result, at least three other techniques were developed and can be used to reduce matrix interference while enabling the use of full scan namely: (a) high-resolution mass spectrometry [3, 4, 5]; (b) GC-MS with Cold electron ionization (EI) in which the enhanced molecular ions suppress matrix interference as described in [6]; (c) two-dimensional comprehensive gas chromatography (GCxGC)-MS with its improved GC separation that can be used for the reduction of matrix interference as described in a few papers with thermal modulation GCxGC-MS [5, 7, 8, 9, 10, 11] and as discussed in this paper.

We note that any GC-MS-based method for improved pesticide analysis should also preferably extend the range of pesticides amenable for GC-MS analysis to include some of the currently “LC-MS only” amenable pesticides and enable improved trace pesticide identification via the NIST library (or other libraries) as well as allow for pesticide identity confirmation by an independent method. Furthermore, this new method should preferably be obtained with a low-cost single quadrupole MS system. In order to meet the above-mentioned challenges, we developed pulsed flow modulation GCxGC-MS with Cold EI method and system [12, 13, 14, 15] for improved pesticide analysis as described in this paper.

GC-MS with Cold EI (EI of cold molecules in supersonic molecular beams (SMB) and fly-through ion source) [16] was evaluated in the past with pesticide analysis in agricultural products as described in [6]. GC-MS with Cold EI provides enhanced molecular ions and it extends the range of pesticides amenable for analysis [6, 16]. Furthermore, the feature of enhanced molecular ions leads to improved selectivity and reduced matrix interference in comparison with lower mass fragments as described in [6]. It was reported that the selectivity of the molecular ion versus matrix interference is improved by about a factor of 20/100 amu higher mass and the molecular ions are the highest mass ions in each compound. We recently repeated some of these experiments and found that the factor of 20 was too high, but nevertheless, on the average, the selectivity of the molecular ions against matrix interference is equivalent to that of MS-MS on fragment ions.

Comprehensive GCxGC was developed by Phillips and co-workers [17, 18, 19] and it is slowly gaining popularity. GCxGC and GCxGC-MS are reviewed in [7, 20, 21]. The most widely used GCxGC modulators are thermal modulators [7, 17, 18, 19]. We developed pulsed flow modulation (PFM) for obtaining GCxGC-MS with SMB (also named Cold EI) based on a low-cost single quadrupole mass analyzer and without any added cooling gas [12, 13, 14, 15]. In PFM-GCxGC-MS with Cold EI, sample compounds which elute from the first GC column are temporarily stored in a fused silica transfer line and are pulsed periodically injected by ~ 25 ml/min He gas pulse into the second column. After the pulse, ~ 20 ml/min He develops the chromatography in the second column for a few seconds (typically 4 s). PFM is simple to construct, does not require any added gas with Cold EI, and the second column injection time can be tuned; hence, PFM-GCxGC-MS with Cold EI is enabled with the limited scan speed of quadrupole MS. The increased PFM second GC column flow rate is seamlessly handled by the Cold EI nozzle vacuum chamber.

We combined PFM-GCxGC with the Aviv Analytical 5975-SMB GC-MS with Cold EI [12] and analyzed with it 12 pesticides in two agricultural matrices. PFM-GCxGC-MS with Cold EI combines the improved separation of GCxGC with the features of Cold EI of enhanced molecular ions and isotope information for the provision of improved sample identification information. Unique to PFM-GCxGC-MS with Cold EI is that in contrast to thermal modulation with standard GC-MS the second-dimension elution time does not suffer from any ion source peak tailing and/or second column saturation effects and thus it is ideally suitable for pesticide analysis in complex matrices. Thus, in GCxGC-MS with Cold EI, the matrix interference is reduced by both the enhanced selectivity of the molecular ion and by the improved separation of GCxGC. The aim of this study is to demonstrate the usefulness of PFM-GCxGC-MS with Cold EI, characterize in a quantitative way its efficiency in the removal of matrix interference, and discuss its use as an alternative to GC-MS-MS.

Pulsed flow modulation (PFM)—the modulator concept of operation

Pulsed flow modulation GCxGC (PFM-GCxGC) and its combination with the 5975-SMB GC-MS with Cold EI (Aviv Analytical, Tel Aviv Israel) is described in details in reference [15]. Thus, it will be described here only as needed for pesticide analysis. Referring to Fig. 1.
Fig. 1

Schematic diagram of pulsed flow modulation for GCxGC-MS

The main idea behind PFM-GCxGC is that by properly tuning flow directions and rates, one can convert the continuous GC output gas flow rate from the first analytical column A into discrete gas pulses for their pulsed injection into the second GCxGC analytical column B. The sample compounds elute from column A into a storage transfer line (typically 50 cm 0.53 mm I.D. deactivated fused silica capillary) through a first T connector. The storage transfer line is connected to the second GCxGC analytical column B via a second T connector. Both T connectors are further connected via a first gas transfer line A and a second gas transfer line B, respectively, to a three-way solenoid valve that is controlled by the PFM operational software, which controls the direction of helium flow. At a given time after the storage transfer line is almost full (typically chosen to be 4 s), a pulse of about 0.3 s of typically 20 ml/min helium is directed by the solenoid valve into the first T connector, and the previously eluted sample compounds which were stored in the storage transfer line are quickly flushed into the second analytical column B while temporarily stopping the first analytical column A flow. After this short gas flow pulse, the three-way solenoid valve changes its direction for a few seconds (in our case, 4 s), the flow in gas transfer line A stops while the flow in the second gas transfer line B commences at a lower flow rate such as 16 ml/min, since transfer line B has a higher flow impedance than transfer line A (the solenoid valve provides constant pressure). This relatively lower flow rate enables the elution and analysis of sample compounds in the second GCxGC analytical column B, while the sample which elutes from the first analytical column A, is stored in the storage transfer line. An important feature of the pulsed flow modulation is that the short, high flow rate pulse, which flushes the sample from the storage transfer line into the second analytical column B, also simultaneously stops the elution of sample compounds from the first analytical column A. An important additional feature of the PFM is that the ~ 20 times higher flow rate in the second GCxGC column B than in the analytical column A focuses and compresses in time the sample compounds that are pulsed injected into the second column and the injection pulse width (in time) can be tuned. For example, when a cycle time of 4 s is used, it is 4 s divided by the flow rate ratio of 20 which results in an injection pulse width of 200 ms. As a result, PFM-GCxGC can be coupled with quadrupole MS despite its limited scan speed. Accordingly, the PFM GCxGC modulator arrangement is simpler than thermal modulation GCxGC modulator and it does not require any cryogenic gas supply with its bulky CO2 cylinders or liquid nitrogen containers.

Experimental

The experimental system

In this section, we outline the full range of experimental parameters used in the pesticides analysis by PFM-GCxGC-MS with Cold EI and explain some of the important considerations behind the selection of central parameters.

System: 5975-SMB GC-MS with Cold EI that is based on the combination of an Agilent 5975 MSD (Agilent Technologies, Santa Clara, CA, USA) with the Aviv Analytical supersonic molecular beam interface and its unique fly-through ion source (Aviv Analytical, Tel Aviv, Israel). A comparison study of the same samples was performed at the Israel Plant Protection center with a standard Agilent 5975 MSD system.

Sample: 12 pesticides as listed in the tables below (“Results” section) that were spiked in baby leaves mixture extract and in lettuce extract.

Injection: 5 μl with pulsed splitless at 70 PSIG for 0.5 min at 220 °C injector temperature.

First-dimension analytical column: 15 m, 0.25 mm ID, 0.25 μ film DB-5ms-UI.

Second-dimension analytical column: 3 m, 0.32 mm ID, 0.25 μ film Rxi-200.

He first column flow rate: 0.8 ml/min.

Pulsed flow modulation gas transfer lines: 70 cm, 0.25 mm ID, capillary transfer line to the second-dimension column entry side and 50 cm, 0.32 mm ID, capillary transfer line to the storage capillary at its first analytical column exit side.

Sample storage capillary: 50 cm of 0.53 mm ID deactivated transfer line by Agilent.

Pulsed flow modulation helium flow rates: about 20 ml/min during the injection time and 16 ml/min during the second column separation cycle time.

GC oven temperature program: 100 °C with 20 °C/min to 150 °C followed by 5 °C/min to 260 °C followed by 20 °C/min to 300 °C with waiting period of 6 min at 300 °C for total run time of 31.5 min.

PFM pressure program: Linear from 2450 mBar at 2.5 min up to 2850 mBar at 25 min.

Backflush: Backflush was initiated after 28.5 min for 3 min via column A flow rate reduction to 0.2 ml/min and the increase of PFM flow rate to 30 ml/min (> 30 PSIG). It resulted in about 3 ml/min backflush flow rate through the first column A and 27 ml/min through column B.

Cold EI fly-through ion source: 10 mA emission, 70 eV electron energy, 62 ml/min total He makeup, and PFM flow rates using 700 mBar nozzle backing pressure. The PFM gas requirement is offset by reduced nozzle make-up gas flow rate thus the PFM by itself does not add any gas requirement.

SMB transfer line temperature: 250 °C with temperature program of 5 °C/min after 20 min to 280 °C.

5975 mass spectrometer: 50–460 amu at 11.6 Hz scanning frequency (0 A/D samples).

5975 mass spectrometer tune change: The faster than usual scan speed introduced negative mass shift of about 0.17 amu that was corrected via the change in the manual tune screen of MASSOFFSET from − 34 to − 24 in the tune values. No other change was made in the tune.

electronic pressure and flow controller: Alicat model MC200 EPC/EFC (Alicat Scientific Tucson AZ USA).

Sample preparation

Sample preparation was based on the Dutch mini-Luke extraction method which is described in reference [22, 23].

Analytical conditions and method optimization considerations

GCxGC method development is known to be more involved than standard GC method development. However, once the PFM-GCxGC concept is understood its method development can be relatively simple to implement and optimize based on the following considerations:
  • First-dimension GC Column: While the main target of PFM GCxGC-MS is the reduction of matrix interference, we also wish to enable the use of the same set-up but without the use of PFM for extended range of pesticides amenable for analysis for the confirmation of LC-MS pesticides. Thus, we selected 0.25 mm ID column with only 15-m length which enables standard operation with 0.8 ml/min with some lower elution temperatures than with the standard 30-m column combined with up to 25 ml/min pulsed splitless injection flow rate at lower than usual injector temperatures. The use of shorter column further enables its operation with up to 15 ml/min flow rate during standard chromatography (without PFM) thereby with lower elution temperatures of labile pesticides. Naturally, if only GCxGC-MS is targeted one can use the standard 30-m columns.

  • Second-dimension GC column: In PFM-GCxGC, it should be as long as possible for best separation and with minimal flow impedance under high column flow rate limits. However, it must enter into 0.53-mm ID storage transfer line hence 0.32-mm ID column is the most suitable. The second column sorption phase selection is the same as in thermal modulation GCxGC and the Restek Rtx-200 is a reasonable choice that induced pesticide separation that covered the full 4-s cycle time of the second GC dimension. The length was chosen to be 3 m which is a tradeoff between minimal flow impedance and separation power.

  • Storage capillary length: This length selection depends on the choice of second-dimension cycle time which is 4 s in our case and first dimension column flow rate which was 0.8 ml/min. In this case, the eluting volume in 4 s is 53 μl and the storage capillary should have volume bigger than that. For 0.53 mm ID transfer line capillary, this means that its net length should be longer than 27 cm. We selected 50 cm to have freedom in the choice of flow rate and cycle time as some added length does not degrade the performance.

  • Pulse duration time and its PFM-GCxGC optimization: We selected 0.32 s. The selection of injection pulse duration and second column flow rate is optimized via the use of a steady flux of pentane from a sampler. We used for PFM pulse width optimization a 1-ml syringe with a ChromatoProbe vial inside it that was loaded with pentane and inserted into the injector to serve as a diffusion source of steady flux of pentane. After ~ 1-min first-column transit time, one can operate the MS with SIM on m/z = 72 of pentane and see the effect of all the PFM parameters on the injection time and pulse shape and width on the Agilent Chemstation screen. The goal is to have injection time of about 200 ms which was obtained in our case with 2450 mBar absolute (~ 21 PSIG) and 0.32-s injection pulse duration. The actually produced pulse width was faster than 0.32 s and was about 0.2 s.

  • PFM gas source lines: In PFM-GCxGC, the gas line that is connected into the first column side must be with smaller flow impedance than the second gas line so that when the flow is switched into it, the pressure at the first column output will be increased and as a result, the first column flow rate will be temporarily halted for the duration of the second-dimension injection time of ~ 0.2 s. We found that the use of 0.32 mm ID capillary for the first column side and 0.25 mm ID capillary for the second-dimension column side are adequate and we even further prolonged the 0.25 mm ID capillary length by about 20 cm.

  • PFM pressure program: Similar to thermal modulation, as the GC oven temperature is increased, the time spread in the second-dimension column is increased and at some GC oven temperature, there is a risk of having second-dimension elution in the next cycle (known as wrap around). Thus, in order to prevent that adverse effect, we used second-column flow program via the second-column head pressure program. Accordingly, the pressure was gradually increased from constant 2450 mBar for 2.5 min, followed by linear pressure program up to 2850 mBar at 25 min and constant pressure of 2850 mBar up to 28.5 min. Between 28.5 - and 31.5-min backflush was operated with 30 ml/min. This pressure program worked very well to eliminate the wrap around problem.

  • GC oven temperature program: The GC oven temperature program is also important for the GCxGC-MS performance. For optimal first-dimension separation, the peak width should be longer than the second-dimension cycle time and up to 3-s dimension cycles. As a result, the temperature program should be slow yet not too slow to provide sufficient analysis throughput. In addition, the GC program also affects the elution time spread in the second dimension. The slower the GC oven temperature program the lower are the sample compounds elution temperatures and thus the greater is their time spread in the second column elution times. Thus, our compromise value was 5 °C/min during the pesticide elution time that resulted in full spread of the pesticide elution times in the second GC dimension combined with standard analysis time of 31.5 min.

  • Backflush: In the analysis of pesticides in complex agricultural matrices, backflush is practically essential for two reasons: (a) It helps to keep the column inert and clean despite the high load of matrix that is deposited on it from the liner; (b) It enables the injection of larger volumes of samples in complex matrices such as up to 5-μl splitless (5 mg matrix extract) while keeping the column in good condition despite the added matrix load on the column. As a result, up to 5 times improved instrument sensitivity can be obtained. One of the advantages of PFM-GCxGC over thermal modulation GCxGC is that the PFM-GCXGC device can also serve as a built-in backflusher device since it includes a gas supply junction with electronic flow/pressure control behind the flow impedance of the second-dimension GCxGC column. Thus, after 28.5 min, the pressure program was switched into backflush mode with 30 ml/min which is about 3000 mBar and in addition, the Agilent 7890 GC injector pressure was reduced to 2 PSIG at that time. Backflush is a known effective technique to keep the first-dimension column clean from low volatility compounds such as agricultural matrices compounds and in our experiments it fully proved its value. The injection of 5-μl splitless was enabled with a standard liner at 70 PSIG since at these conditions the volume of vaporized acetone solvent is still smaller than the liner volume.

  • 5975 scan speed: We used 10.6 Hz that, as shown in the “Results” section below, is sufficient for the pesticide analysis with PFM-GCxGC-MS with Cold EI. For the mass spectral range of 50–460 amu, this is a scanning rate of 4400 amu/s. The Agilent 5975 MSD is capable of using faster scan mode up to about 8000 amu/s. However, we found that when the fast scan mode is used, isotope abundance information is lost since the mass spectral resolution is deteriorated. Since we believe that GCxGC-MS can benefit from the provision of best sample identification information, we prefer to have somewhat slower than maximal scan speed than losing the ability to provide pesticide identification confirmation with our TAMI software [19] for improved sample identification based on molecular ion isotopomers height ratio.

Results

PFM-GCxGC-MS with cold EI analysis of pesticides in agricultural matrices

We evaluated PFM-GCxGC-MS with Cold EI in the analysis of 12 pesticides that were spiked at 100 ppb (ng/g produce) and at 25 ppb (ng/g produce) in baby leaves mixture extract and in lettuce extract. The baby leaves mixture is considered as a relatively “difficult matrix” while the lettuce is an easier matrix. In Fig. 2 below we show a comparison of pesticide analysis by standard Agilent 5975 GC-MS with its standard EI ion source (upper trace) and by PFM-GCxGC-MS with Cold EI (bottom trace). 0.3-min sections of the total ion count mass chromatograms are shown in which 100 ng/g methidathion elutes at 11.94 min in standard EI and at 14.51 min in PFM-GCxGC-MS with Cold EI. As demonstrated, the methidathion pesticide is revealed in standard EI only as a small peak at the shoulder of a much higher matrix peak and both are superimposed on some matrix background. In sharp contrast, in PFM-GCxGC-MS with Cold EI, the methidathion is revealed as an almost clean narrow peak with little matrix interference. Furthermore, the Cold EI mass spectrum of methidathion is characterized by significant enhancement of the molecular ion. Thus, the combination of enhanced molecular ion and largely reduced matrix interference results in “dramatic” improvement in this pesticide analysis by PFM GCxGC-MS with Cold EI in both peak to matrix background ratio and in the quality of its mass spectrum that is far cleaner with PFM-GCxGC-MS with Cold EI.
Fig. 2

A comparison of pesticide analysis by standard GC-MS with its standard EI (upper traces) and PFM-GCxGC-MS with Cold EI (bottom traces). 0.3 min sections of the total ion count mass chromatograms are shown in which 100 ng/g methidathion (in baby leaves mixture extract) elutes below the indicating arrows at 11.94 min in standard EI and at 14.51 min in PFM-GCxGC-MS with Cold EI. Note the vast improvement obtained in PFM GCxGC-MS with Cold EI in both peak to matrix background ratio and in the quality of mass spectrum that is far cleaner with PFM-GCxGC-MS with Cold EI. The PFM TIC peak at 14.44 min is also of methidathion in an earlier GCxGC cycle

In order to have a quantitative measure of the magnitude of improvement that is obtained by the PFM-GCxGC in the reduction of matrix interference, we need to establish a measurable factor of improvement. The best way that we found to quantify this improvement factor is by the ratio of pesticide peak to matrix background and thus the PFM improvement factor is the ratio of the pesticide peak to matrix in PFM-GCxGC-MS with Cold EI divided by the same ratio as obtained in GC-MS with standard EI. In PFM-GCxGC-MS with Cold EI, the methidathion/matrix ratio is 12.0 while in GC-MS with standard EI, it is 0.077 and thus the PFM improvement factor is 155. In fact, it is even higher than 155 since standard EI was operated with 30-m column while PFM-GCxGC-MS with Cold EI was operated with 15-m column thus with about 1.4 times less first-dimension separation.

We found that with all the 12 pesticides (described in Table 1 below), PFM-GCxGC-MS with Cold EI helped to significantly reduce the matrix interference in comparison with GC-MS with standard EI. The PFM improvement factors ranged from 3 up to 155 and the average PFM improvement factor is 36. This high magnitude of improvement needs an explanation about its origin. Clearly PFM GCxGC improves the GC separation hence reduces the matrix interference. However, the second-dimension GCxGC cycle time is 4 s and average pesticide peak width is 250–300 μs and thus the expected improvement factor is about 14 and considering the fact that each pesticide and matrix compound span about two GCxGC cycles time means that the theoretically expected PFM improvement factor is only 7 at most.
Table 1

PFM-GCxGC-MS with Cold EI gain in matrix interference reduction (PFM improvement factors) in the analysis of 12 pesticides at 100 ng/g concentration versus (a) GC-MS with standard EI with baby leaves mixture matrix, (b) GC-MS with Cold EI with baby leaves mixture matrix, and (c) GC-MS with Cold EI with lettuce matrix

Pesticide name

Molecular weight

Elemental formula

PFM gain factor vs GC-MS with SEI baby leaves

PFM gain factor vs GC-MS with CEI baby leaves

PFM gain factor vs GC-MS with CEI lettuce

Dimethoate

229.00

C5H12NO3PS2

63

48

33

Ethoxyquin

217.15

C14H19NO

16

13

69

Diazinon

304.10

C12H21N2O3PS

25

23

30

Chlorpyrifos

348.93

C9H11Cl3NO3PS

3

>100

10

Methidation

301.96

C6H11N2O4PS3

155

41

15

Mepanipyrim

223.11

C14H13N3

3

20

>100

Tebuconazole

307.15

C16H22ClN3O

8

3

21

Iprodione

329.03

C13H13Cl2N3O3

~100

18

14

Bifenthrin

422.13

C23H22ClF3O2

14

14

18

Bifenazate

300.15

C17H20N2O3

16

13

6

Pyriproxyfen

321.14

C20H19NO3

9

8

26

Prochloraz

375.03

C15H16Cl3N3O2

25

20

52

This expected factor is also based on the assumption that the pesticides and matrix elution time in the second dimension is statistically (randomly) distributed and fully and uniformly fill the 4-s second-dimension elution time windows. However, fortunately, pesticides are all semi-polar compounds while agricultural matrices are mostly non-polar or of little polarity and up to high polarity in free fatty acids. In Fig. 3, we show the two-dimensional plot of PFM-GCxGC-MS with Cold EI data of the 12 pesticides mixture spiked at 100 ng/g in baby leaves mixture extract (1 g/ml). The pesticides elution times are indicated by their names and arrows to their elution peaks. As shown, the pesticides are all mid polarity hence are mostly in second dimension GCxGC elution times that are with relatively little matrix. Thus, while usually the second-dimension elution times are not fully orthogonal to that of the first-dimension column, in PFM-GCxGC-MS with Cold EI, we encounter a rare case of “Super Orthogonality” meaning that the separation of the second dimension is better than expected based on its cycle time and average pesticide peak widths.
Fig. 3

A two-dimensional plot of PFM-GCxGC-MS with Cold EI analysis data of 12 pesticides mixture spiked at 100 ng/g in baby leaves mixture extract (1 g/ml)

It is emphasized that the large improvement obtained in the rejection of matrix interference is predominantly due to the improved GC separation of PFM-GCxGC-MS and only marginally due to the Cold EI ion source features. Thus, we similarly encountered significant PFM improvement factors in the comparison of PFM-GCxGC-MS with Cold EI with GC-MS with Cold EI. Figure 4 shows a comparison of pesticide analysis by GC-MS with Cold EI (upper trace) and PFM-GCxGC-MS with Cold EI (bottom trace). 0.4 min sections of the total ion count mass chromatograms are shown in which 100 ng/g iprodione (in baby leaves mixture 1 g/ml extract) elutes at the indicating time arrows at 21.76 min in GC-MS with Cold EI and 21.11 min in PFM-GCxGC-MS with Cold EI. As demonstrated, significant improvement is obtained with PFM-GCxGC-MS with Cold EI in both peak to matrix background ratio and in the quality of iprodione mass spectra that is far cleaner with PFM-GCxGC-MS with Cold EI.
Fig. 4

A comparison of pesticide analysis by GC-MS with Cold EI (upper trace) and PFM-GCxGC-MS with Cold EI (bottom trace). The 0.6- and 0.4-min sections of the total ion count mass chromatograms are shown in which 100 ng/g iprodione (in baby leaves mixture extract) elutes at the indicating time arrows at 21.76 min in GC-MS with Cold EI and 21.11 min in PFM-GCxGC-MS with Cold EI. Note the significant improvement obtained in PFM-GCxGC-MS with Cold EI in both peak to matrix background ratio and in the quality of mass spectra that is far cleaner with PFM-GCxGC-MS with Cold EI

In Table 1, we summarize the PFM improvement factors in the analysis of 12 pesticides at 100 ppb (ng/g) concentration versus: (a) GC-MS with standard EI with baby leaves mixture matrix, (b) GC-MS with Cold EI with baby leaves mixture matrix, and (c) GC-MS with Cold EI with lettuce matrix. As described in Table 1, the PFM improvement factors are high for all the three comparisons and they span the range of 3 up to over 100. The average PFM improvement factors are 36 versus standard GC-MS in baby leaved mixture, 27 versus GC-MS with Cold EI in baby leaves mixture, and 33 versus GC-MS with Cold EI in lettuce. As a result, the overall averaged PFM improvement factor is determined as 32.

This average PFM improvement factor of 32 is considered by us as very high, higher even from the average matrix interference reduction factor of GC-MS-MS versus GC-MS which we estimate as 20. Furthermore, it is also much higher than the general impression and perception of average improvement in GCxGC-MS with TOF and standard EI [8]. The reasons for the apparent superiority of PFM-GCxGC-MS with Cold EI over GCxGC-MS with thermal modulation TOF-MS and standard EI are (A) Cold EI excels in having ultra-fast ion source response time and no ion source peak tailing even for the most polar sample and/or matrix compounds and (B) PFM-GCxGC-MS is operated with 0.32 mm ID large-diameter second column with high (20 ml/min) second column flow rate while thermal modulation GCxGC is typically operated with a micro bore 0.1 mm ID second column and with standard 1 ml/min flow rate. As a result, the sample capacity of the second column in PFM-GCxGC is higher by over two orders of magnitude than in thermal modulation.

More importantly, while in thermal modulation, the second-dimension column is easily saturated and exhibits large peak fronting that hampers the GCxGC separation; in PFM-GCxGC, the second column never exhibits second-column saturation even if the first column is saturated by the matrix. The topic of second GCxGC column peak fronting is discussed in ref. [24].

An interesting example that demonstrates both advantages of PFM-GCxGC-MS with Cold EI over GCxGC-MS with standard EI is shown in Fig. 5 that shows the case of full first-column coelution of the polar matrix compound free palmitic acid and the pesticide chlorpyrifos. As demonstrated in Fig. 5, chlorpyrifos is almost fully separated by the PFM-GCxGC and as a result, it provides a very clean and informative Cold EI mass spectrum for its effective identification and quantitation. However, the importance of Fig. 5 is far greater than of a demonstration of a case of effective PFM-GCxGC separation. In Fig. 5, we also demonstrate that (A) A very polar-free fatty acid saturates the first-dimension GC column that exhibits clear peak fronting while highly symmetric peaks of this matrix compound are exhibited in the second-dimension GCxGC mass chromatogram. Thus, Fig. 5 proves the claim that unlike in thermal modulation GCxGC in PFM-GCxGC the second column never saturates even if the first dimension column saturates and is fronting; (B) Clean PFM-GCxGC-MS with Cold EI peaks are observed without any tailing for both the chlorpyrifos pesticide and for the highly polar-free palmitic fatty acid. This is in sharp contrast with any GC-MS with standard EI results in which free fatty acids require derivatization for their analysis.
Fig. 5

A section of PFM-GCxGC-MS with Cold EI full scan mass chromatogram of 100 ng/g pesticides in baby leaves mixture extract. The mass chromatogram is zoomed at the 12.10–12.55 min elution time window. As shown, chlorpyrifos (indicated by the arrow) is fully separated by the PFM-GCxGC from the much larger amount of free palmitic acid (n-hexadecanoic acid C16H32O2) that coeluted in the first-dimension GC column and saturated it. As a result, high-quality Cold EI-MS of Chlorpyrifos is shown at the bottom trace (after minor background subtraction) and it resulted in 97% NIST library identification probability

In Fig. 6, we demonstrate the extreme peak tailing that can occur with polar matrix compounds inside the standard EI ion source. Figure 6 shows the 10–20-min elution time section of GC-MS with standard EI reconstructed mass chromatogram at m/z = 256.1 of baby leaves mixture extract. As shown, the free Palmitic acid eluted as a relatively small peak with very long peak tail of over 10 min.
Fig. 6

A 10–20 min elution time section of GC-MS with standard EI reconstructed mass chromatogram at m/z = 256.1 of baby leaves mixture extract. As shown, the free palmitic acid eluted as a relatively small peak with very long peak tail of over 10 min

Peak tailing hampers the effectiveness of GCxGC-MS with standard EI ion sources in the analysis of pesticides in agricultural matrices for two main reasons: (A) polar pesticides can severely tail hence lose their added GCxGC separation and (B) highly polar matrix compound such as free fatty acids can tail and impede the GCxGC-MS separation. As shown in Fig. 5, PFMxGC-GC-MS with Cold EI does not suffer from any ion source peak tailing even for free fatty acids. Thus, the combination of absence of ion source peak tailing and second-column peak fronting contributes to and result in the high effectiveness of PFM-GCxGC-MS with Cold EI and in its high average PFM improvement factor of 32 in the analysis of pesticides in agricultural products.

While the usual emphasis in pesticide analysis is on their quantitative determination, their positive identification is also highly valuable as the impact of false negative and/or false positive identification of pesticides can be severe. In fact, false identification of pesticides in agricultural products can lead to major loss of money and food that is thrown away. Clearly, NIST library identification is the best tool for pesticide identification. In comparison with MS-MS, it also includes the step of exclusion of over 270,000 compounds in addition to the identification of the specific pesticide while in MS-MS, the identification is based only on matching of the MS-MS peaks to the pesticide without excluding other possible compounds. However, the problem with library identification is that it is often severely impeded by matrix interference.

In Table 2 below we show that, at the concentration level of 25 ppb (ng/g produce), none of the 12 pesticides were identified by the NIST library in baby leaves mixture extract in GC-MS with standard EI (no background subtraction was attempted). In fact, only 10 out of the 12 pesticides were included in our NIST 2005 library but we trust that newer versions of the NIST library include many more compounds and, in any case, pesticides that are not included can be added by the user to the library. In sharp contrast, in PFM-GCxGC-MS with Cold EI, 9 out of the 10 pesticides that are in the NIST library where identified and all of them were number one in the NIST hit list and with reasonable identification probabilities. Thus, the high PFM improvement factors manifest themselves also in highly improved NIST identification probabilities. We note that contrary to some perception Cold EI with its enhanced molecular ions actually improves the identification probability with the NIST library as demonstrated and discussed in details in [25].
Table 2

Pesticides identification at 25 ng/g concentration in baby leaves mixture extract

Name

Average Mass

Elemental Formula

TAMI SIM ions

TAMI Match in SIM

NIST prob.

(PFM CEI)

NIST prob.

(GC-SEI)

Dimethoate

229.00

C5H12NO3PS2

229.0, 231.0

851

45.9

NI

Ethoxyquin

217.15

C14H19NO

217.1, 218.1

763

82.9

NI

Diazinon

304.10

C12H21N2O3PS

304.1–306.1

991

87.2

NI

Chlorpyrifos

348.93

C9H11Cl3NO3PS

348.9, 350.9, 352.9

999

51.5

NI

Methidation

301.96

C6H11N2O4PS3

302.0, 304.0

999

92.4

NI

Mepanipyrim

223.11

C14H13N3

222.1223.1, 224.1

NC

NA

NA

Tebuconazole

307.15

C16H22ClN3O

307.1–309.1

857

49.1

NI

Iprodione

329.03

C13H13Cl2N3O3

329.0, 331.0

998

72.2

NI

Bifenthrin

422.13

C23H22ClF3O2

422.1–424.1

903

54.0

NI

Bifenazate

300.15

C17H20N2O3

300.1, 301.1

975

NA

NA

Pyriproxyfen

321.14

C20H19NO3

321.1322.1

729

49.4

NI

Prochloraz

375.03

C15H16Cl3N3O2

375, 377, 379

974

NI

NI

Column four shows the ions that were selected for identification in SIM mode by the Tal-Aviv Molecule Identifier software. Column five shows the obtained TAMI matching factors for the pesticides (0–1000 when 1000 denotes a perfect match). Column six shows the obtained NIST identification probabilities in full-scan PFM-GCxGC-MS with Cold EI in the analysis of 12 pesticides at 25 ng/g concentration in baby leaves mixture. Column seven shows the obtained NIST identification probabilities in full scan GC-MS with standard EI in the analysis of 12 pesticides at 25 ng/g concentration in baby leaves mixture

SEI standard EI, CEI Cold EI, NI not identified, NA not applicable such as not in the NIST library, and NC not confirmed meaning TAMI match below 700

Once a pesticide was identified its identification (and quantification) should be confirmed by an independent method. We developed a method and software for such independent identification which we named Tal-Aviv Molecule Identifier software. TAMI software operation is based on isotope abundance analysis (IAA) [26]. This software automatically confirms or rejects NIST library identification, and the combined use of two independent identification methods, the NIST library and TAMI-based provision of elemental formula provides unambiguous identification. If NIST library fails (such as when the compound is not in the library), the TAMI software still provides elemental formula for identification, and the available EI-MS fragments provides additional structural information. However, effective use of the TAMI software requires good isotopomer ion ratios, which is often not available in the analysis of trace-level pesticides. Clearly the large PFM improvement factors are helpful in the reduction of possible matrix interference.

Thus, our solution for the performance of independent pesticide confirmation identification was the analysis of time programmed few ions monitoring in which the SIM ions were two or three molecular ion isotopomers. The selected time programmed SIM ions are listed in Table 2 at its column four. As shown in Table 2 column 5, the identification of 11 out of the 12 pesticides were confirmed by the TAMI software with matching factors above 700 and 9 with matching factors above 800. Accordingly, PFM-GCxGC-MS with Cold EI excels not only in improved identification of pesticides by the NIST library but also by the confirmation of the pesticide identification with the TAMI software which requires having sufficiently abundant molecular ions as provided by Cold EI.

Discussion

Pulsed flow modulation GCxGC-MS with Cold EI uniquely combines improved GCxGC separation with the superior MS selectivity and sample identification power of Cold EI. Furthermore, PFM-GCxGC is easier to construct, maintain, and operate than thermal modulation GCxGC. An additional important advantage of PFM-GCxGC is that it is fully compatible with the scan speed of quadrupole mass analyzers in view of its ability to tune the second GC dimension injection time, while with Cold EI, its increased ~ 20–30 ml/min flow rate is seamlessly handled without any use of open split that can lead to reduced sensitivity. As a result, PFM-GCxGC-MS with Cold EI can provide a superior combination of improved separation and identification at lower cost of goods compared to thermal modulation GCxGC combination with TOF-MS with standard EI.

This article describes the application of PFM-GCxGC-MS with Cold EI for pesticides analysis in agricultural matrices. Twelve pesticides were explored at the concentration levels of 100 and 25 ppb (ng/g produce) in both baby leaves mixture extract and in lettuce extract. For improving the instrumental sensitivity, we injected splitless 5 μl of extracts (5 mg commodity) and in order to keep the first dimension GC column in good condition backflush was applied at the end of each run. We note that backflush capability is inherently provided with PFM-GCxGC.

Very high PFM improvement factors were obtained and the average matrix interference reduction factor was 32. This large average PFM improvement factor implies that PFM-GCxGC-MS with Cold EI is as effective or even more effective than GC-MS-MS in the removal of matrix interference. Furthermore, PFM-GCxGC-MS with Cold EI operates in the full scan mode and thus enables significantly improved pesticide identification by the NIST library combined with an independent pesticide identification confirmation. This confirmation is achieved via their molecular ions isotope ratios using time programmed two or three ions SIM and the Aviv Analytical TAMI software to obtain machine generated matching factors that serve for an independent pesticide identification confirmation.

PFM-GCxGC-MS with Cold EI was demonstrated in this paper as an effective method for pesticides analysis in agricultural products. Based on this study we conclude the following:
  1. 1.

    No ion source tailing and/or second column fronting. PFM-GCxGC-MS with Cold EI unlike GCxGC-MS with standard EI is characterized by the full elimination of any ion source peak tailing and/or by the absence of second GCxGC column saturation effects (peak fronting) that plague thermal modulation GCxGC-MS with standard EI.

     
  2. 2.

    Enhanced molecular ions and its related improved selectivity. Cold EI provides enhanced molecular ions for further improved pesticide quantitation and suppression of matrix interference that is much lower on high mass ions. The combination of added selectivity of Cold EI on the molecular ion and its isotopomers plus the reduction of matrix interference by the improved PFM-GCxGC separation makes PFM-GCxGC-MS with Cold EI far superior to GC-MS-MS on fragment ions.

     
  3. 3.

    Largely reduced matrix interference. PFM-GCxGC-MS with Cold EI provided average PFM-GCxGC improvement factor of 32 that is equivalent or better than the matrix interference reduction of MS-MS. Thus, PFM-GCxGC-MS with Cold EI can serve as an alternative to the widely used GC-MS-MS with similar or superior effectiveness in matrix interference reduction.

     
  4. 4.

    An alternative to GC-MS-MS. An important question is if PFM-GCxGC-MS with Cold EI is similar to the established GC-MS-MS in the removal of matrix interference, why should it serve as an alternative to GC-MS-MS. The answer to this question is simple. GC-MS-MS is a fully target-oriented analysis method and any pesticide that is not included in its limited list of analyzed pesticides is prone to false negative identification [10]. Accordingly, most of the pesticides are actually excluded from GC-MS-MS analysis and are simply ignored. Furthermore, MS-MS must include compound-specific calibrations and its related large stock of pesticide sample solutions since its response is highly compound-dependent as it depends on the relative abundance of the selected parent ion as well as on the efficiency of the CID in the provision of daughter ions which can significantly vary from pesticide to pesticide (and system to system). In addition, as the number of pesticides in the MS-MS list is growing, its instrumental sensitivity is reduced, and in a list of about 300 pesticides, about 10–30 pesticides can be included in a single elution time window and lead into signal reduction almost at the same magnitude as in full scan in view of the fact that the signal is shared by up to 60 ions and suffer from about 10 times signal losses in the MS-MS process.

     
  5. 5.

    Universal pesticide analysis—the main goal. In contrast to GC-MS-MS, PFM-GCxGC-MS with Cold EI is operated in the full scan mode and thus its generated files can serve any time for post run data analysis of unlimited number of screened pesticides. This data analysis goal is achieved via reconstructed single or few ions monitoring (RSIM) on the pesticides molecular ions plus additional isotopomers or high mass fragments from an unlimited list of pesticides in predetermined elution time windows. It should be emphasized that Cold EI is characterized by uniform compound independent response in its total ion mass chromatogram peaks, the same as in standard EI but in standard EI unlike in Cold EI, the response uniformity is eroded by ion source tailing for large and polar pesticides. Thus, while a selected group of few hundreds “important” pesticides (those that are selected for MS-MS) can be analyzed with calibration curves, all the other several hundred pesticides can be included in the list of RSIM analysis in given known elution time windows and listed molecular ions plus additional isotopomers or high mass fragments. The semi-quantitative determination of these added pesticides can be performed even without the use of calibration pesticide solutions and it can be accurate within a factor of two via the TIC peak area or via RSIM on the molecular ion that is calibrated to the total ion peak. While the assumed accuracy this way is only a factor of two, it is much better than to ignore the majority of all pesticides and have zero accuracy in their measurements. Once a new pesticide is consistently found, it can be included in the list of the “important” pesticides for its higher accuracy analysis.

     
  6. 6.

    Improved pesticides identification and confirmation. PFM-GCxGC-MS with Cold EI uniquely enables largely improved pesticide identification with the NIST library even at trace levels in complex matrices. Furthermore, it allows for the performance of independent pesticides identification confirmation via the measurement of their molecular ions isotopomers ratios and the generation of unbiased machine generated matching factors with the TAMI software.

     
  7. 7.

    Extended range of pesticides amenable for analysis and the confirmation of LC-MS pesticides. The same PFM-GCxGC-MS with Cold EI system can also uniquely serve for the independent confirmation of LC-MS pesticides. It can be used for this purpose with high first-column flow rate without the pulsing of the PFM that will only add in this mode a few milliliters per minute flow rate to the second column. With 15-m first-column length, the injection flow rate can be over 25 ml/min and column flow rate can be 15 ml/min for achieving significant reduction in the elution temperatures of labile pesticides from both the injector into the column and from the column to the SMB. The use of a temperature programmable injector is preferred for this purpose. Any pesticide decomposition at the Cold EI fly-through ion source ion is fully eliminated. In fact, even in this paper, the feature of extended range was proven via the analysis of free palmitic acid and thus both labile and more polar pesticides that require LC-MS for their analysis can be confirmed with the PFM-GCxGC-MS with Cold EI system.

     

In this paper, we discussed PFM-GCxGC-MS with Cold EI while using a simple single quadrupole MS. An interesting topic to discuss is the suitability of PFM-GCxGC-MS with Cold EI to other higher-end MS systems. Naturally, PFM-GCxGC-MS with Cold EI can be coupled with a Q-TOF and/or high-resolution TOF with the added benefits of further improved selectivity in view of the higher MS resolution and ability to perform MS-MS (with Q-TOF) for independent confirmation. However, an important benefit of the use of a single quadrupole MS is its simplicity, ease of operation, robustness, and low cost and thus we do not think that there is a need for PFM-GCxGC-MS with Cold EI combination with a Q-TOF or Orbitrap. With Orbitrap, we predict that the use of its C trap may lead into undesirable dissociation of the enhanced molecular ions while its scan speed is further limited compared to a quadrupole MS thus while the Orbitrap is a great MS for pesticide analysis, it is not a good marriage with PFM-GCxGC-MS with Cold EI.

Conclusions

We believe that PFM-GCxGC-MS can serve as a viable alternative to the industry-standard GC-MS-MS. We estimate the MS-MS matrix interference reduction factor in GC-MS to be about 20 times versus the four ions RSIM method, and it lowers the LOD by about an order of magnitude. It is important to note that from this estimated factor of 20 reduced matrix interference by MS-MS, the inherent MS-MS selectivity is only about a factor of 5–6 while additional gain factor of 3–4 comes from the fact that full-scan RSIM (or timed shared SIM) employs four ions and the matrix interference on the weaker and lower mass ion numbers 3 and 4 are particularly strong while MS-MS is performed only on two parent ions that are typically of higher mass. With the use of Cold EI, we believe that in view of availability of the molecular ions, three ions that includes the molecular ions are enough for superior pesticides identification over four fragment ions. The molecular ion is unique in its compound identification capability as by itself it eliminates false identification of homologous and degradation compounds that share the same fragments. Our suggested approach for pesticide identification begins with NIST library identification which is vastly improved with PFM-GCxGC-MS with Cold EI as demonstrated and discussed above and if such identification failed we use the molecular ion, one abundant isotopomer such as M + 1 or M + 2 and one high mass fragment ion. The measured ratio of isotopomer/molecular ion is another powerful pesticide identification tool. In terms of instrument sensitivity PFM-GCxGC-MS with Cold EI can be used with time programmed three ions monitoring for certain list of selected “important” pesticides, the same as in MS-MS, but simultaneously with full scan for the monitoring of unlimited number of pesticides.

We claim that the selectivity of PFM-GCxGC-MS against matrix interference is superior to that of GC-MS-MS since the GCxGC separation alone improves the separation against the matrix by an average factor of 32 which is bigger than the average matrix reduction factor of MS-MS plus in addition, Cold EI by itself improves the matrix interference suppression via its feature of enhanced molecular ions since the reduced matrix interference on the molecular ion is equivalent to that of MS-MS on a fragment ion. Thus, the combination of PFM-GCxGC and Cold EI is superior to MS-MS in matrix interference reduction. Similarly, the instrumental sensitivity of Cold EI in SIM mode is superior to that of MS-MS in view of inherent MS-MS losses in its Q2, CID process and Q3 (about a factor of ×5–10). Furthermore, while not demonstrated in this article, Cold EI enables the analysis of larger and more thermally labile “LC-MS” pesticides [27]. PFM-GCxGC-MS with Cold EI suffer from two downsides of harder quantitation in view of having a few PFM signal pulses that need to be integrated (summed), and its speed of analysis is typically longer although in this paper we used “standard” pesticides analysis time and with further loss of selectivity the analysis time can be reduced to be below 10 min. Accordingly, while Figs. 1 and 3 serve as good demonstrations of the advantages and benefits of PFM-GCxGC and Cold EI method, the multiple “split” peak nature of GCxGC should be pointed out and a software should be devised to add the peaks with the same spectra together. However, we consider the ability to perform universal pesticide screening analysis without any limitation on the number of screened pesticide as the most important attribute of PFM-GCxGC-MS with Cold EI and thus it can and should serve as a superior alternative to GC-MS-MS.

Notes

Funding

This research was supported by the Israel Science Foundation founded by the Israel Academy of Sciences and Humanities (grant No 356/15). This research was also supported by the Ministry of Science and Technology, Israel, and by the Pazi grant number 266/15 of the Israel Atomic Energy Commission.

Compliance with ethical standards

Conflict of interest

The authors declare that they have no competing interests.

References

  1. 1.
    Turner JA. Pesticides manual. 17th edition. British Crop Protection Council; 2015.Google Scholar
  2. 2.
    Pesticide Manual Online. 2017. http://www.bcpc.org/product/bcpc-online-pesticide-manual-latest-version. Accessed 20 Mar 2017.
  3. 3.
    Mol HGJ, Tienstra M, Zomer P. Evaluation of gas chromatography-electron ionization-full scan high resolution Orbitrap mass spectrometry for pesticide residue analysis. Anal Chim Acta. 2016;935:161–72.CrossRefPubMedGoogle Scholar
  4. 4.
    Xia D, Gao L, Zheng M, Wang S, Liu G. Simultaneous analysis of polychlorinated biphenyls and polychlorinated naphthalenes by isotope dilution comprehensive two dimensional gas chromatography high-resolution time-of-flight mass spectrometry. Anal Chim Acta. 2016;937:160–7.CrossRefPubMedGoogle Scholar
  5. 5.
    Ochiai N, Ieda T, Sasamoto K, Takazawa Y, Hashimoto S, Fushimi A, et al. Stir bar sorptive extraction and comprehensive two-dimensional gas chromatography coupled to high-resolution time-of-flight mass spectrometry for ultra-trace analysis of organochlorine pesticides in river water. J Chromatogr A. 2011;1218:6851–60.CrossRefPubMedGoogle Scholar
  6. 6.
    Kochman M, Gordin A, Goldshlag P, Lehotay SJ, Amirav A. Fast, high-sensitivity, multipesticide analysis of complex mixtures with supersonic gas chromatography-mass spectrometry. J Chromatogr A. 2002;974:185–212.CrossRefPubMedGoogle Scholar
  7. 7.
    Mondello L, Tranchida PQ, Dugo P, Dugo G. Comprehensive two-dimensional gas chromatography-mass spectrometry: a review. Mass Spectrom Rev. 2008;27:101–24.CrossRefPubMedGoogle Scholar
  8. 8.
    Hoh E, Lehotay SJ, Pangallo KC, Mastovska K, Ngo HL, Reddy CM, et al. Simultaneous quantitation of multiple classes of organohalogen compounds in fish oils with direct sample introduction comprehensive two-dimensional gas chromatography and time-of-flight mass spectrometry. J Agric Food Chem. 2009;57:2653–60.CrossRefPubMedGoogle Scholar
  9. 9.
    Hoh E, Lehotay SJ, Mastovska K, Ngo HL, Vetter W, Pangallo KC. Capabilities of direct sample introduction-comprehensive two-dimensional gas chromatography-time-of-flight mass spectrometry to analyze organic Chemicals of Interest in fish oils. Environ Sci Technol. 2009;43:3240–7.CrossRefPubMedGoogle Scholar
  10. 10.
    Hoh E, Dodder NG, Lehotay SJ, Pangallo KC, Reddy CM, Maruya KA. Nontargeted comprehensive two-dimensional gas chromatography/time-of-flight mass spectrometry method and software for inventorying persistent and bioaccumulative contaminants in marine environments. Environ Sci Technol. 2012;46:8001–8.CrossRefPubMedGoogle Scholar
  11. 11.
    Fernandes VC, Lehotay SJ, Geis-Asteggiante L, Kwon H, Mol HGJ, van der Kamp H, et al. Analysis of pesticide residues in strawberries and soils by GC-MS/MS, LC-MS/MS and two dimensional GC-time-of-flight MS comparing organic and integrated pest management farming. Food Addit Contam Part A. 2014;31:262–70.CrossRefGoogle Scholar
  12. 12.
    Poliak M, Kochman M, Amirav A. Pulsed flow modulation comprehensive two-dimensional gas chromatography. J Chromatogr A. 2008;1186:189–95.CrossRefPubMedGoogle Scholar
  13. 13.
    Poliak M, Fialkov AB, Amirav A. Pulsed flow modulation two-dimensional comprehensive gas chromatography-tandem mass spectrometry with supersonic molecular beams. J Chromatogr A. 2008;1210:108–14.CrossRefPubMedGoogle Scholar
  14. 14.
    Amirav A. Pulsed flow modulation gas chromatography mass spectrometry with supersonic molecular beams method and apparatus. Israel Patent application number 176724 (2006), USA Patent 7518103 submitted in 2007 issued 2009.Google Scholar
  15. 15.
    Keshet U, Fialkov AB, Alon T, Amirav A. A New pulsed flow modulation GC×GC–MS with Cold EI system and its application for jet fuel analysis. Chromatographia. 2016;79:741–54.Google Scholar
  16. 16.
    Amirav A, Gordin A, Poliak M, Fialkov AB. Gas chromatography-mass spectrometry with supersonic molecular beams. J Mass Spectrom. 2008;43:141–63.CrossRefPubMedGoogle Scholar
  17. 17.
    Liu Z, Phillips JB. Comprehensive two-dimensional gas chromatography using an on-column thermal modulator interface. J Chromatogr Sci. 1991;29:227–31.CrossRefGoogle Scholar
  18. 18.
    Phillips JB, Xu J. Comprehensive Multi-dimensional gas chromatography. J Chromatogr A. 1995;703:327–34.Google Scholar
  19. 19.
    Phillips JB, Ledford EB. Thermal modulation: a chemical instrumentation component of potential value in improving portability. Field Anal Chem Technol. 1996;1:23–9.CrossRefGoogle Scholar
  20. 20.
    Dalluge J, Beens J, Brinkman UAT. Comprehensive two-dimensional gas chromatography: a powerful and versatile analytical tool. J Chromatogr A. 2003;1000:69–108.CrossRefPubMedGoogle Scholar
  21. 21.
    Mondello L. Comprehensive chromatography in combination with mass spectrometry. Wiley; 2011.Google Scholar
  22. 22.
    Dutch mini-Luke extraction method “Analytical Methods for Pesticide Residues in Foodstuffs”, 6th edition. The Hague: Inspectorate for Health Protection, Ministry of Health, Welfare and Sports; 1996.Google Scholar
  23. 23.
    Lozano A, Kiedrowska B, Scholten J, de Kroon M, de Kok A, Fernández-Alba AR. Miniaturisation and optimisation of the Dutch mini-Luke extraction method for implementation in the routine multi-residue analysis of pesticides in fruits and vegetables. Food Chem. 2016;192:668–81.CrossRefPubMedGoogle Scholar
  24. 24.
    Harynuk J, Gorecki T, de Zeeuw J. Overloading of the second-dimension column in comprehensive two-dimensional gas chromatography. J Chromatogr A. 2005;1071:21–7.CrossRefPubMedGoogle Scholar
  25. 25.
    Alon T, Amirav A. How enhanced molecular ions in Cold EI improve compound identification by the NIST library. Rapid Commun Mass Spectrom. 2015;29:2287–92.Google Scholar
  26. 26.
    Alon T, Amirav A. Isotope abundance analysis methods and software for improved sample identification with supersonic gas chromatography/mass spectrometry. Rapid Commun Mass Spectrom. 2006;20:2579–88.CrossRefPubMedGoogle Scholar
  27. 27.
    Fialkov AB, Gordin A, Amirav A. Extending the range of compounds amenable for gas chromatography–mass spectrometric analysis. J Chromatogr A. 2003;991:217–40.CrossRefPubMedGoogle Scholar

Copyright information

© Springer-Verlag GmbH Germany, part of Springer Nature 2017

Authors and Affiliations

  1. 1.School of ChemistryTel Aviv UniversityTel AvivIsrael
  2. 2.Israel Plant Protection and Inspection ServicesBeit DaganIsrael

Personalised recommendations