Intraoperative awareness due to malfunction of a Siemens 900B ventilator

  • 212 Accesses

  • 4 Citations

Abstract

A case of intraoperative awareness during a thoracotomy is described. The patient’s recall coincided with an intraoperative period during which a Siemens 900B ventilator and a Siemens 952 isoflurane vaporiser were used. Subsequent assessment of this equipment with an anaesthetic agent analyzer revealed that, at the ventilator settings which had been used, the delivered anaesthetic vapour concentration varied greatly from the vaporizer settings. This problem eventually was traced to a malfunctioning inlet control valve on the ventilator. This complication may have been prevented if the end-tidal anaesthetic concentration had been monitored intraoperatively.

Résumé

Nous décrivons un cas d’anesthesie insuffisiante non détectée, avec souvenirs des evenements perioperatoire, dûe a une valve d’entrée sur un ventilateur Siemens 900B ne fonctionnant pas. Ce problème aurait pu étre éviter en monitorant la concentration dagent volatile dans les gazs expirés.

References

  1. 1

    Siemens Isoflurane vaporizer 952 operating manual: Sth ed, Siemens-Elena AB (Solna, Sweden) 1985: 4.

  2. 2

    Gross JB, Alexander CM. Awakening concentrations of isoflurane are not affected by analgesic doses of morphine. Anesth Analg 1988; 67: 27–30.

  3. 3

    Guerra F. Awareness and recall. Int Anesthcsiol Clin 1986; 24(4): 75–99.

  4. 4

    Schultetus RR, Hill CR, Dharamraj CM, Banner TE, Berman LS. Wakefulness during Cesarean section after anesthetic induction with ketamine, thiopental or ketamine and thiopental combined. Anesth Analg 1986; 65: 723–8.

  5. 5

    Aldrete JA. Is the patient asleep? Int Surg 1987; 72: 58–61.

  6. 6

    Mori K. The EEG and awareness during anaesthesia. Anaesthesia 1987; 42: 1153–5.

Download references

Author information

Correspondence to Peter D. Slinger.

Rights and permissions

Reprints and Permissions

About this article

Cite this article

Slinger, P.D., Andrew, W., Scott, C. et al. Intraoperative awareness due to malfunction of a Siemens 900B ventilator. Can J Anaesth 37, 258–261 (1990). https://doi.org/10.1007/BF03005481

Download citation

Key words

  • anaesthesia: depth, awareness
  • complications: awareness
  • equipment: ventilators, malfunction