Qualitative und quantitative Fluoreszenzmikrospektrographie mit dem Leitz-Mikrospektrographen

Qualitative and quantitative fluorescence cytophotometry at the Leitz-Microspectrograph

  • 15 Accesses

Summary

The use and additional equipment of the Leitz-Mikrospectrograph for microspectrofluorometry and fluorescence cytophotometry are described in detail.

Fluorescence excitation with incident light together with the use of dichromatic beam splitters for optimum separation of the excitation beam from emitted radiation are the characteristic features of this instrument.

The fluorescence light is spectrally dispersed by a glass-prism (Special glass for visible light, quarz for UV).

A motor driven oscillating mirror moves along the spectrum and reflects the light of continuously changing wavelength on to the photo-cathode of the multiplier, that transfers the optical impulse into an electrical signal. The wavelength dependend change of spectral intensity is registered by a light beam oscillograph fitted with four fixed potentiometers indicating the 400, 500, 600 and 700 nm wavelength position by crosspoints with an oblique line drawn by the wavelength potentiometer, while the sixth potentiometer records the spectral curve.

Problems of spectral calibration and correction are discussed.

When the prism and the oscillating mirror are removed the instrument may be used as fluorescence cytophotometer. For this purpose of quantitative fluorescence cytophotometry a digital measuring unit with a print-out instrument is presented.

Zusammenfassung

Ausgehend vom Leitz-Mikrospektrographen wird gezeigt, wie dieses Gerät für mikrospektrofluorimetrische Untersuchungen und für die quantitative Fluoreszenzcytophotometrie ausgerüstet werden kann. Die im Zusammenhang mit der Anwendung des Gerätes auftretenden Fragen der Eichung und Korrektur der Spektralkurven werden erörtert.

This is a preview of subscription content, log in to check access.

Literatur

  1. Björklund, A., Ehinger, B., Falck, B.: A method for differentiating dopamine in tissue sections by microspectrofluorometry. J. Histochem. Cytochem. 16, 263–270 (1968).

  2. Böhm, N., Sprenger, E.: Fluorescence cytophotometry: A valuable method for the quantitative determination of nuclear Feulgen-DNA. Histochemie 16, 100–118 (1968).

  3. —: “Nine-position-test” for proof of even illumination of the light field in fluorescence microspectrophotometry. Histochemie 22, 59–61 (1970).

  4. Brumberg, E. M., Krylova, T. N.: O fluoreschentnykh mikroskopakh. Zh. obshch. biol. 14, 461 (1953).

  5. —, Meissel, M. N., Grigorjeva, T. A., Barsky, I. J., Gutkina, A. V.: Ultraviolet fluorescence microscopy as a new field of histochemistry. Ann. Histochim., Suppl. 2, 135–142 (1962).

  6. Grehn, J., Kornmann, H.: Kontrastfluoreszenz mit Opak-Illuminator. Leitz-Mitt. Wiss. u. Techn. 3, Nr 4, 108–111 (1965).

  7. Kemplay, J. R.: The electronic requirements of a spectrofluorimeter. Electron. engng. 34, 820–823 (1962).

  8. Lippert, E., Nägele, W., Seibold-Blankenstein, I., Staiger, U., Voss, W.: Messung von Fluoreszenzspektren mit Hilfe von Spektralphotometern und Vergleichsstandards. Z. analyt. Chem. 170, 1–17 (1959).

  9. Loeser, Ch. N., West, S. S.: Cytochemical studies and quantitative television fluorescence and absorption spectroscopy. Ann. N. Y. Acad. Sci. 97, (2), 346–357 (1962).

  10. Pearse, A. G. E., Rost, F. W. D.: A microspectrofluorimeter with epi-illumination and photon counting. J. Microscopy 89, (3), 321–328 (1969).

  11. Ploem, J. S.: Die Möglichkeit der Auflichtfluoreszenzmethoden bei Untersuchungen von Zellen in Durchströmungskammern und Leighton-Röhren. Acta biochem. (Jena) 7, 339–342 (1967a).

  12. —: The use of vertical illuminator with interchangeable dichroic mirrors for fluorescence microscopy with incident light. Z. wiss. Mikr. 68, 129–142 (1967b).

  13. —: Ein neuer Illuminator-Typ für die Auflicht-Fluoreszenzmikroskopie. Leitz-Mitt. Wiss. u. Techn. 4, Nr 8, 225–238 (1969).

  14. Ruch, F.: Ein Mikrospektrograph für Absorptionsmessungen im ultravioletten Licht. Z. wiss. Mikr. 64, 453–468 (1960).

  15. Sprenger, E., Böhm, N.: Eine Möglichkeit automatisierter Meßwerterfassung in der Fluoreszenzcytophotometrie. Acta histochem. (Jena), Suppl. 10 (im Druck).

  16. Thaer, A.: Speicherbetrieb mit einem Superorthicon in der Mikrospektrographie. Haus der Technik-Vortragsveröffentlichungen. „Fernsehen in der Industrie“ 69, 49–55 (1966).

Download references

Author information

Rights and permissions

Reprints and Permissions

About this article

Cite this article

Sprenger, E., Böhm, N. Qualitative und quantitative Fluoreszenzmikrospektrographie mit dem Leitz-Mikrospektrographen. Histochemie 25, 163–176 (1971). https://doi.org/10.1007/BF00279114

Download citation