Decelerating inspiratory flow waveform improves lung mechanics and gas exchange in patients on intermittent positive-pressure ventilation

  • 114 Accesses

  • 100 Citations

Abstract

The effects of two inspiratory flow waveforms (WFs), decelerating and constant have been studied in 14 patients undergoing intermittent positive-pressure ventilation (IPPV). With tidal volume (VT), inspiratory time, inspiratory-expiratory (I/E) ratio and frequency being kept constant, the decelerating waveform produced statistically significant reduction of peak pressure, total respiratory resistance, work of inspiration, ratio of dead space to tidal volume (VD/VT) and alveolar-arterial gradient for oxygen (A-a)PO2. There was significant increase in total static and kinetic compliances and PaO2, with no significant changes in PaCO2, in cardiac output (CO) and in other haemodynamic measurements.

This is a preview of subscription content, log in to check access.

References

  1. 1.

    Adams AP, Economides AP, finley WEI, Sykes MK (1970) The effects of variations of inspiratory flow waveform on cardiorespiratory function during controlled ventilation in normo-, hypo- and hypervolaemic dogs. Br J Anaesth 42:818

  2. 2.

    Baker AB, Babington PCB, Colliss JE, Cowie RW (1977) Effects of varying inspiratory flow waveform and time in intermittent positive pressure ventilation. I. Introduction and methods. Br J Anaesth 49:1207

  3. 3.

    Baker AB, Colliss JE, Cowie RW (1977) Effects of varying inspiratory flow waveform and time in intermittent positive pressure ventilation. II. Various physiological variables. Br J Anaesth 49:1233

  4. 4.

    Bergman NA (1963) Effect of different pressure breathing patterns on alveolar-arterial gradients in dogs. J Appl Physiol 18:1049

  5. 5.

    Bergman NA (1967) Effects of varying inspiratory waveforms on gas exchange. Anesthesiol 28:390

  6. 6.

    Bergman NA (1969) Effect of varying respiratory waveforms on distribution of inspired gas during artificial ventilation. Am Rev Respir Dis 100:518

  7. 7.

    Chackrabarti MK, Sykes MK (1976) Evaluation of the lung ventilator performance analyser. Anaesthesia 31:521

  8. 8.

    Cox LA, Almeida AP, Robinson JS, Horsley JK (1974) An electronic respirometer. Br J Anaesth 46:302

  9. 9.

    Cox LA, Chapman EDW (1974) A comprehensive volume cycled lung ventilator embodying feedback control. Med Biol Eng 12:160

  10. 10.

    Don H (1977) The mechanical properties of the espiratory system during anasthesia. Int Anaesthesiol Clin 15(2):113

  11. 11.

    Don HF, Robson JG (1965) The mechanics of the respiratory system during anesthesia. Anesthesiol 26(2):168

  12. 12.

    Enghoff H (1938) Volumen Inefficax Bemerkungen zur Frage des schadlichen Raumes. Upsala Lak Foren Forh 44:191

  13. 13.

    Fahri LE (1966) Ventilation-perfusion relationships and its role in alveolar gas exchange. In: Collin G Care (ed) Advances in Respiratory Physiology. Edward Arnold London, p 178

  14. 14.

    Fairley HB (1970) Artificial ventilation. In: Scurr C, Feldman S (eds) Scientific foundations of anaesthesia. William Heinemann, London, p 192

  15. 15.

    Fuleihan SF, Wilson RS, Pontoppidan H (1976) Effect of mechanical ventilation with end-inspiratory pause on blood-gas exchange. Anesth Analg 55(1):122

  16. 16.

    Herzog P, Norlander OP (1968) Distribution of alveolar volumes with different types of positive pressure gas flow patterns. Opuscula Medica 13:27

  17. 17.

    Hughes R, May AJ, Widdicombe JG (1959) Stress relaxation in rabbit's lungs. J Physiol (Lond) 146:85

  18. 18.

    Ingelstedt S, Jonson B, Nordstrom L, Olsson S-G (1972) A servo-controlled ventilator measuring expired minute volume, airway flow and pressure. Acta Anaesth Scand (Suppl) 47:7

  19. 19.

    Johansson H, Löfström JB (1975) Effects on breathing mechanics and gas exchange of different inspiratory gas flow patterns during anaesthesia. Acta Anaesth Scand 19:8

  20. 20.

    Johansson H (1975a) Effects on breathing mechanics and gas exchange of different inspiratory gas flow patterns in patients undergoing respirator treatment. Acta Anaesth Scand 19:19

  21. 21.

    Johansson H (1975b) Effects of different inspiratory gas flow patterns on thoracic compliance during respirator treatment. Acta Anaesth Scand 19:89

  22. 22.

    Lyager S (1968) Influence of flow pattern on the distribution of the respiratory air during intermittent positive-pressure ventilation. Acta Anaesth Scand 12:191

  23. 23.

    Mushin WW, Rendell-Baker L, Thompson PW, Mapleson WW (1980) Automatic ventilation of the lung. Third edn, Blackwell, Oxford

  24. 24.

    Nunn JF (1977) Applied respiratory physiology, Butterworths, London

  25. 25.

    Nunn JF, Bergman NA, Coleman AJ (1965) Factors influencing the arterial oxygen tension during anaesthesia with artificial ventilation. Br J Anaesth 37:898

  26. 26.

    Otis AB, Mckerrow CB, Bartlett RA, Mead J, Mellroy MB, Selverstone NJ, Radford EP (1956) Mechanical factors in distribution of pulmonary ventilation. J Appl Physiol 8:427

  27. 27.

    Otis AB (1964) Quantitative relationship in steady-state gas exchange. In: Fenn O, Rahn H (eds) Handbook of physiology, section 3. Respiration, vol 1. American Physiological Society, Washington D.C., p 681

  28. 28.

    Robinson JS (1978) Respiratory care and the principle of ventilation. In: Hanson GC, Wright PL (eds) The medical management of the critically ill. Academic Press, London, p 5

  29. 29.

    Robinson JS (1973) The influence of differing flow patterns on the gas distribution and uptake during artificial ventilation. 7th International Anaesthesia Postgraduate Course. H. Egermann, Vienna, p 277

  30. 30.

    Watson WE (1962a) Some observations on dynamic lung compliance during intermittent positive pressure respiration. Br J Anaesth 34:153

  31. 31.

    Watson WE (1962b) Transmural central venous pressure during intermittent positive pressure respiration. Br J Anaesth 34:278

  32. 32.

    Watson WE (1962c) Observations on physiological dead space during intermittent positive pressure respiration. Br J Anaesth 34:502

  33. 33.

    Watson WE (1962d) Observations on the dynamic lung compliance of patients with respiratory weakness receiving intermittent positive pressure respiration. Br J Anaesth 34:690

  34. 34.

    Watson WE, Spalding JMK, Smith AC (1962) A respirator pump to provide a variable waveform. Br J Anaesth 34:149

Download references

Author information

Rights and permissions

Reprints and Permissions

About this article

Cite this article

Al-Saady, N., Bennett, E.D. Decelerating inspiratory flow waveform improves lung mechanics and gas exchange in patients on intermittent positive-pressure ventilation. Intensive Care Med 11, 68–75 (1985). https://doi.org/10.1007/BF00254777

Download citation

Key words

  • IPPV
  • Inspiratory WFs
  • VD/VT ratio
  • PaO2
  • (A-a)PO2 and CO