Abstract

Purpose

A new self-expanding tulip sheath was tested for percutaneous wire capturing in a femoro-femoral pull-through maneuver.

Methods

In six sheep, a self-expanding tulip sheath was introduced into the right iliac artery. A 5 Fr catheter and an 0.035-inch guidewire were guided into the iliac artery from a left femoral approach. Thirty-six tests were performed to guide the guidewire out of the right groin using the tulip sheath, by catching the catheter first in 18 tests and just by catching the guidewire in the other 18. As a control, six tests were performed using a conventional sheath. We performed one clinical arterial application.

Results

There was prompt capture of the wire in 35 of 36 tests using a self-expanding sheath. Five of six tests using a conventional sheath failed. The clinical application of the tulip principle was successful.

Conclusion

Using the self-expanding tulip sheath to facilitate wire capture during pull-through maneuvers, even in difficult situations, is promising.

This is a preview of subscription content, log in to check access.

References

  1. 1.

    Loose H, Ryall T (1988) Common iliac artery occlusion: Treatment with pull-through angioplasty. Radiology 168:273–274

  2. 2.

    Gaines P, Cumberland D (1988) Wire loop technique for angioplasty of total iliac artery occlusions. Radiology 168:275–276

  3. 3.

    Vorwerk D, Guenther RW (1990) Mechanical revascularization of occluded iliac arteries with use of self-expandable endoprostheses. Radiology 175:411–415

  4. 4.

    Fogarty TJ, Cranley JJ, Krause RJ, Strasser ES, Hafner CD (1963) A method of extraction of arterial emboli and thrombi. Surg Gynecol Obstet 116:241–244

  5. 5.

    Vorwerk D, Guenther RW, Clerc C, Schmitz-Rode T, Imbert C (1992) Percutaneous embolectomy: In-vitro investigations of the self-expanding tulip sheath. Radiology 182:415–418

Download references

Author information

Rights and permissions

Reprints and Permissions

About this article

Cite this article

Vorwerk, D., Günther, R.W., Schürmann, K. et al. New device for wire and catheter capturing: In vivo experiments and first clinical experience. Cardiovasc Intervent Radiol 18, 312–314 (1995). https://doi.org/10.1007/BF00203682

Download citation

Key words

  • Cross-over technique
  • Wire loop technique
  • Pull-through maneuver
  • Tulip sheath
  • Self-expanding sheath