Abstract
The performance and applicability of the hand-held Raman spectrometer FirstDefender RM (Thermo Scientific) to nuclear safeguards field were assessed. More than 100 different uranium ore concentrates, frequently referred to as yellow cake, were measured and wherever applicable library scans were collected. A subset of the samples was then used to test the identification capability of the instrument. The accuracy of identification was shown to be sound for uranyl peroxide, sodium diuranate, ammonium diuranate, uranium trioxide and ammonium uranyl carbonate. The study also demonstrated the instrument’s potential for a quick quantitative determination of the uranium concentration in uranyl nitrate solutions.
Similar content being viewed by others
References
FirstDefender RM User manual. Ahura Scientific, Part 110-00013-05
Matthews RL, Ong KY, Brown CD, Zhu L, Knopp K (2006) Evaluation of Ahura’s FirstDefender chemical identifier, Edgewood Chemical Biological Center. U.S. Army Research, Development and Engineering Command, Technical Report ECBC-TR-461
Matthews RL, Longworth TL, Ong KY, Zhu L, Brown CD (2006) Testing of Ahura’s FirstDefender chemical identifier against toxic industrial chemicals. Edgewood Chemical Biological Center, U.S. Army Research, Development and Engineering Command, Technical Report ECBC-TR-528
The structure and content of agreements between the agency and states required in connection with the Treaty of the Non-Proliferation of Nuclear Weapons. INFCIRC/153 (Corrected), IAEA, Reprinted in November 2010
Model protocol additional to the agreement(s) between state(s) and the International Atomic Energy Agency for the application of safeguards. INFCIRC/540 (Corrected), IAEA, Reprinted in September 2011
Lewis IR, Edwards HGM (eds) (2001) Handbook of Raman spectroscopy: from the research laboratory to the process line. Marcel Dekker, Inc, New York
Das RS, Agrawal YK (2011) Raman spectroscopy: recent advancements, techniques and applications. Vib Spectrosc 57:163
BWTek Inc. http://bwtek.com/technology/raman/. Accessed 02 Dec 2014
Agiltron Inc. http://www.agiltron.com/raman/english/pinpointer.htm. Accessed 02 Dec 2014
Rigaku Raman Technologies. http://www.rigakuraman.com/products/. Accessed 02 Dec 2014
Thermo Scientific. http://www.ahurascientific.com/chemical-explosives-id/products/index.php. Accessed 02 Dec 2014
SciAps, Inc. http://sciaps.com/portable-raman-spectrometers/. Accessed 04 April 2015
Snowy Range Instruments. http://wysri.com/cbex. Accessed 02 Dec 2014
Gnyba M, Smulko J, Kwiatkowski A, Wierzba P (2011) Portable Raman spectrometer-design rules and applications. Bull Pol Acad Sci 59(3):325–329
Bastians S, Crump G, Griffith WP, Withhall R (2004) Raspite and studtite: Raman spectra of two unique minerals. J Raman Spectrosc 35:726–731
Volkovich VA, Griffiths TR, Fray DJ, Fields M (1998) Vibrational spectra of alkali metal (Li, Na and K) uranates and consequent assignment of urinate ion site symmetry. Vib Spectrosc 17:83–91
Cordfunke EHP (1970) Composition and structure of ammonium uranates. J Inorg Nucl Chem 32:3129
Stuart WI, Miller DJ (1973) The nature of ammonium uranates. J Inorg Nucl Chem 35:2109
Urbanek V, Sara V, Moravec J (1979) Study of formation and composition of ammonium uranates. J Inorg Nucl Chem 41:537
Ho Mer Lin D, Manara D, Varga Z, Berlizov A, Fanghänel T, Mayer K (2013) Applicability of Raman spectroscopy as a tool in nuclear forensics for analysis of uranium ore concentrates. Radiochim Acta 101:779–784
Armstrong DP, Jarabek RJ, Fletcher WH (1989) Micro-Raman spectroscopy of selected solid UxOyFz compounds. Appl Spectrosc 43:461–468
Anderson A, Chieh C, Irish DE, Tong JPK (1980) An X-ray crystallographic, Raman and infrared spectral study of crystalline potassium uranyl carbonate K4UO2(CO3)3. Can J Chem 58:1651
Pointurier F, Marie O (2010) Identification of the chemical forms of uranium compounds in micrometer-size particles by means of micro-Raman spectrometry and scanning electron microscope. Spectrochim Acta Part B 65:797
Ho MLD, Manara D, Reis PL, Fanghänel T, Mayer K (2014) The use of different dispersive Raman spectrometers for the analysis of uranium compounds. Vib Spectrosc 73:102–110
Palacios LM, Taylor ST (2000) Characterization of uranium oxides using in situ micro-Raman spectroscopy. Appl Spectrosc 54:1372–1378
Khulbe PK, Agarwal A, Raghuvanshi GS, Bist HD, Hashimoto H, Kitagawa T, Little TS, Durig JR (1989) Raman studies of the vibrational dynamics and phase transitions in uranyl nitrate hexahydrate. J Raman Spectrosc 20:283–290
Volod’ko LV, Sabilo KV (1966) Raman spectra of solutions of uranyl compounds. Zh Prikl Spektrosk 4:52–57
Acknowledgments
This work was done under Task A 00860 of the European Commission’s Support Programme to the IAEA. The authors would like to thank Nexia Solutions for providing the UOC samples from their Springfield plant. Also the Canadian Support Programme and the Australian Support Programme to the IAEA shall be recognized for providing UOC samples.
Author information
Authors and Affiliations
Corresponding author
Electronic supplementary material
Below is the link to the electronic supplementary material.
Rights and permissions
About this article
Cite this article
Berlizov, A., Ho, D.M.L., Nicholl, A. et al. Assessing hand-held Raman spectrometer FirstDefender RM for nuclear safeguards applications. J Radioanal Nucl Chem 307, 285–295 (2016). https://doi.org/10.1007/s10967-015-4160-z
Received:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10967-015-4160-z