To the Editor,
We thank Dr. Henlin et al.1 for their interest in our work2 and for sharing their interesting insights into the “parallel intubation technique.” We are pleased that the authors’ experience is consistent with our previously published study results2,3 suggesting that glottis visualization with the Vie Scope® (Adroit Surgical LLC, Oklahoma City, OK, USA) is superior to direct laryngoscopy and noninferior to Macintosh videolaryngoscopy, despite increased time-to-intubation.2,3 In our opinion, the good visualization achieved with the Vie Scope is primarily due to direct epiglottis lifting, being a key element of straight-blade techniques and also significantly improving the glottis exposure in difficult Macintosh videolaryngoscopy.4
Nevertheless, some potential disadvantages of the Vie Scope must be considered:
-
1)
The Vie Scope is designed to be used in a two-step bougie-assisted approach that prolongs time-to-intubation compared with a direct approach.2,3
-
2)
The tracheal tube is railroaded over the bougie without visualization of the laryngeal inlet; therefore, tube advancement cannot be visualized and tube impingement on the anterior commissure or arytenoids cannot be visually confirmed.3
-
3)
Correct bougie placement may not be verified through the Vie Scope in situations with significantly restricted glottis view through the channel (i.e., grade 2C or worse);4,5 in our previous study, we observed four unintended esophageal intubations in patients with expected difficult airways2 probably due to bougie misplacement.
The “parallel intubation technique” suggested by Henlin et al. is very promising for avoiding “blind insertion” of the tracheal tube over the bougie and could furthermore identify a bougie misplacement before the tracheal tube is advanced. Nevertheless, this method retains a two-step approach with an inherently longer time-to-intubation. It must be considered that this technique changes the alignment between bougie and tracheal inlet; therefore, the straight bougie supplied by the manufacturer might not be suitable and a J-shaped bougie may be more appropriate. Nevertheless, this technique is not outlined by the manufacturer and no data are available so far. In our opinion, a study is required to investigate the feasibility, safety, and efficiency of this novel approach.
Thus, there are currently four described options for using the Vie Scope: 1) the median approach, 2) the paraglossal approach, 3) Vie Scope-guided tube placement (after bougie placement and reinsertion of the Vie Scope), and 4) the parallel approach, suggested by Henlin et al. In our opinion, all these approaches should be further investigated to enable a more context-dependent and personalized use of the Vie Scope tailored to the individual patient and specific clinical situation.
References
Henlin T, Michalek P, Gaszynski T, Soltes J. Parallel intubation technique with the Vie Scope® laryngoscope. Can J Anesth 2024; https://doi.org/10.1007/s12630-023-02683-0
Petzoldt M, Grün C, Wünsch VA, Bauer M, Hardel TT, Grensemann J. Vie Scope® versus videolaryngoscopy in expected difficult airways: a randomized controlled trial. Can J Anesth 2023; 70: 1486–94. https://doi.org/10.1007/s12630-023-02534-y
Petzoldt M, Engels Y, Popal Z, et al. Elective tracheal intubation with the Vie Scope®—a prospective randomized non-inferiority pilot study (VieScOP-Trial). Front Med 2022; 9: 820847. https://doi.org/10.3389/fmed.2022.820847
Wünsch VA, Köhl V, Breitfeld P, et al. Hyperangulated blades or direct epiglottis lifting to optimize glottis exposure in difficult Macintosh videolaryngoscopy: a non-inferiority analysis of a prospective observational study. Front Med (Lausanne) 2023; 10: https://doi.org/10.3389/fmed.2023.1292056
Kohse EK, Siebert HK, Sasu PB, et al. A model to predict difficult airway alerts after videolaryngoscopy in adults with anticipated difficult airways—the VIDIAC score. Anaesthesia 2022; 77: 1089–96. https://doi.org/10.1111/anae.15841
Disclosures
Martin Petzoldt has received a study grant from Verathon Inc., Bothell, WA, USA. Tim T. Hardel has received research support from Drägerwerk; Jörn Grensemann has received research support from Ambu and ETView, and consultant fees from Drägerwerk and GE HealthCare. The other authors declare no conflicts of interests.
Funding statement
Open Access funding enabled and organized by Projekt DEAL. All expenses were covered from departmental resources.
Editorial responsibility
This submission was handled by Dr. Stephan K. W. Schwarz, Editor-in-Chief, Canadian Journal of Anesthesia/Journal canadien d’anesthésie.
Author information
Authors and Affiliations
Corresponding author
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License, which permits any non-commercial use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by-nc/4.0/.
About this article
Cite this article
Petzoldt, M., Grün, C., Wünsch, V.A. et al. In reply: Parallel intubation technique with the Vie Scope® laryngoscope. Can J Anesth/J Can Anesth 71, 437–438 (2024). https://doi.org/10.1007/s12630-023-02685-y
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s12630-023-02685-y