Abstract
Interventional cardiologists face significant radiation exposure during interventional cardiology procedures. Therefore, this study focuses on assessing radiation exposure among interventional cardiologists during their procedures. Specifically, it aims to determine the effectiveness of both single and double dosimeter methods in estimating annual occupational radiation doses. This research holds pioneering significance as it represents the very first study undertaken in Sri Lanka. Thirteen interventional cardiologists performed 486 interventional cardiology procedures over three months in three different healthcare institutes. Active Hp(10) dosimeters were placed to measure radiation exposure. Effective doses were calculated using single and double dosimetric algorithms. Annual occupational doses were assessed on an operator basis. Statistical analyses were conducted to assess algorithmic differences and dose variations using the Kruskal–Wallis test and linear regression. The highest annual occupational dose for each dosimetric algorithm received as 2.00 ± 0.24 mSv, 2.29 ± 0.48 mSv, 3.35 ± 0.71 mSv, and 2.64 ± 0.42 mSv, respectively, and remained below the recommended safety limit of 20 mSv/year. The Kruskal–Wallis test revealed no significant differences in the effective doses among double dosimetric algorithms, as well as between single and double dosimetric algorithms (p > 0.05). Linear regression showed strong correlations among various algorithms, demonstrating consistency. The findings of this study hold significant effects on interventional cardiology practice in Sri Lanka, enhancing radiation safety and monitoring.
Similar content being viewed by others
References
http://www.statistics.gov.lk/. The Department of Census and Statistics, Sri Lanka. 2023. 2023. Accessed 4 Oct 2023.
Sri Lanka College of Cardiology [Internet]. 2023. Available from: https://slcc.lk/, URL: https://slcc.lk/. Accessed 4 Oct 2023.
Moladoust H, Ghazanfari-Tehran M, Nikseresht V, Nadim T, Rad MA. Comparison of five developed algorithms to estimate staff effective dose in interventional cardiology: Are they interchangeable? J Med Imaging Heal Informatics. 2015;5:647–51. https://doi.org/10.1166/jmihi.2015.1438.
Biso SMR, Vidovich MI. Radiation protection in the cardiac catheterization laboratory. J Thorac Dis. 2020;12:1648–55. https://doi.org/10.21037/jtd.2019.12.86.
Sowby FD. Annals of the ICRP. Ann ICRP. 1981;6:1. https://doi.org/10.1016/0146-6453(81)90127-5.
López PO, Dauer LT, Loose R, Martin CJ, Miller DL, Vañó E, Doruff M, Padovani R, Massera G, Yoder C. ICRP Publication 139: Occupational Radiological Protection in Interventional Procedures. Ann ICRP. 2018;47:1–118. https://doi.org/10.1177/0146645317750356.
Martin CJ. Personnel dosimetry in UK radiology: Is it time for a change? J Radiol Prot. 2012;32:2–6. https://doi.org/10.1088/0952-4746/32/1/E03.
Kuipers G, Velders XL, De Winter RJ, Reekers JA, Piek JJ. Evaluation of the occupational doses of interventional radiologists. Cardiovasc Intervent Radiol. 2008;31:483–9. https://doi.org/10.1007/s00270-008-9307-7.
Järvinen H, Buls N, Clerinx P, Miljanić S, Nikodemová D, Ranogajec-komor M, Struelens L, d’Errico F. Comparison of double dosimetry algorithms for estimating the effective dose in occupational dosimetry of interventional radiology staff. Radiat Prot Dosimetry. 2008;131:80–6. https://doi.org/10.1093/rpd/ncn239.
Martin CJ. Personal dosimetry for interventional operators: when and how should monitoring be done? Br J Radiol. 2011;84:639–48. https://doi.org/10.1259/bjr/24828606.
Kuipers G, Velders XL. Effective dose to staff from interventional procedures: estimations from single and double dosimetry. Radiat Prot Dosimetry. 2009;136:95–100. https://doi.org/10.1093/rpd/ncp155.
Schultz FW, Zoetelief J. Estimating effective dose for a cardiac catheterisation procedure with single or double personal dosemeters. Radiat Prot Dosimetry. 2006;118:196–204. https://doi.org/10.1093/rpd/ncl018.
Padovani R, Foti C, Malisan MR. Staff dosimetry protocols in interventional radiology. Radiat Prot Dosimetry. 2001;94:193–6. https://doi.org/10.1093/oxfordjournals.rpd.a006471.
Niklason LT, Victoria Marx M, Chan HP. The estimation of occupational effective dose in diagnostic radiology with two dosimeters. Health Phys. 1994;67:611–5. https://doi.org/10.1097/00004032-199412000-00003.
The Swiss Federal Department of the Interior and the Federal Department of the Environment, Transport E and C. Swiss Ordinance for personal dosimetry (Dosimetry Regulation). 1994, Swiss Ordinance for personal dosimetry, edited by the Federal Chancellery, Bern 1999.
Baechler S, Gardon M, Bochud F, Sans-Merce M, Trueb Ph, Verdun FR. Personnel dosimetry in fluoroscopy. Paris: Proc. of Second European IRPA Congress, (Société Françarse de Radioprotection) P-091, 1–10, 2006.
Von Boetticher H, Lachmund J, Hoffmann W. An analytic approach to double dosimetry algorithms in occupational dosimetry using energy dependent organ dose conversion coefficients. Health Phys. 2010;99:800–5. https://doi.org/10.1097/HP.0b013e3181e850da.
Kim KP, Miller DL, Balter S, Kleinerman RA, Linet MS, Kwon D, Simon SL. Occupational radiation doses to operators performing cardiac catheterization procedures. Health Phys. 2008;103:80–99. https://doi.org/10.1097/HP.0b013e31824dae76.
Fardid R, Mirzadeh F, Rezaei H. Occupational doses of cardiologists in cath labs and simulation method. J Cancer Res Ther. 2017;13:901–7. https://doi.org/10.4103/0973-1482.192767.Cited:in: (PMID: 29237950).
Paulsen GU. Assessment of effective dose and dose to the lens. Radiat Prot Dosimetry. 2008;132:313–8.
Aalbers T, Duane S, Kapsch R-P, Meghzifene A. Measurement uncertainty—a practical guide for secondary standards dosimetry laboratories. 2008. URL: https://www-pub.iaea.org/MTCD/publications/PDF/te_1585_web.pdf. Accessed 4 Oct 2023.
Trianni A, Padovani R, Foti C, Cragnolini E, Chizzola G, Toh H, Bernardi G, Proclemer A. Dose to cardiologists in haemodynamic and electrophysiology cardiac interventional procedures. Radiat Prot Dosimetry. 2006;117:111–5. https://doi.org/10.1093/rpd/nci764.
Ingwersen M, Drabik A, Kulka U, Oestreicher U, Fricke S, Krankenberg H, Schwencke C, Mathey D. Physicians’ radiation exposure in the catheterization lab: Does the type of procedure matter? JACC Cardiovasc Interv. 2013;6:1095–102. https://doi.org/10.1016/j.jcin.2013.05.012.
Kaljevic J, Ciraj-Bjelac O, Stankovic J, Arandjic D, Bozovic P, Antic V. Occupational dose assessment in interventional cardiology in Serbia. Radiat Prot Dosimetry. 2016;170:279–83. https://doi.org/10.1093/rpd/ncv439.
Bahruddin NA, Hashim S, Karim MKA, Sabarudin A, Ang WC, Salehhon N, Bakar KA. Radiation dose to physicians’ eye lens during interventional radiology. J Phys Conf Ser. 2016. https://doi.org/10.1088/1742-6596/694/1/012035.
Alemayehu TG, Bogale GG, Bazie GW. Occupational radiation exposure dose and associated factors among radiology personnel in Eastern Amhara, Ethiopia. PLoS One [Internet]. 2023;18:1–14. https://doi.org/10.1371/journal.pone.0286400.
Kim J, Seo S, Lee DN, Park S, Im KJ, Park S, Jin YW. Occupational exposure characteristics and factors associated with radiation doses among Korean Radiation Workers. Radiat Prot Dosimetry. 2020;189:106–13. https://doi.org/10.1093/rpd/ncaa019.
Dalah EZ, Mahdi O, Elshami W, Abuzaid MM, David LR, Mira OA, Obaideen A, Elmahdi HM, Bradley DA. Occupational doses to cardiologists performing fluoroscopically-guided procedures. Radiat Phys Chem [Internet]. 2018;153:21–6. https://doi.org/10.1016/j.radphyschem.2018.09.008.
Najafi M, Nedaie HA, Lahooti A, Omranipour R, Nafissi N, Akbari ME, Olfatbakhsh A, Kaviani A, Alavi N. Radiation exposure of the surgeons in sentinel lymph node biopsy. Iran J Radiat Res. 2012;10:53–7.
Acknowledgements
This work was supported by the General Sir John Kotelawala Defence University, Sri Lanka [KDU/RG/2021/FAHS/003]. We declare no conflicts of interest. Sachini Udara Wickramasinghe was involved in the conceptualization, data extraction, data analysis, and drafting of the manuscript. Vijitha Ramanathan and Sivananthan Sarasanandarajah were involved in conceptualization, checking the accuracy of data extraction and analysis, reviewing, and supervision.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
All the authors have declared no conflict of interest. The study obtained institutional ethical clearance and informed consent from all the participants.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
About this article
Cite this article
Wickramasinghe, S.U., Ramanathan, V. & Sarasanandarajah, S. Assessment of dosimetric approaches in evaluating radiation exposure for interventional cardiologists in Sri Lanka. Radiol Phys Technol 17, 258–268 (2024). https://doi.org/10.1007/s12194-023-00774-7
Received:
Revised:
Accepted:
Published: